Accutom-50
Instruction Manual
4
Always remember to switch the power off when installing electrical
equipment!
The factory setting for Accutom-50 is 240V.
Pull out the fuse holder from the cable terminal at the back of the
machine.
Pull out the voltage switch and turn it to the correct setting.
Voltage Required
Setting
230 or 240V
240V
200 to 220V
200V
Note: The two additional settings, 110V and 120V are not to be used.
Reinsert the voltage switch.
Put the fuse holder back into the cable terminal.
Supplying Power
DANGER!
The machine must be earthed
IMPORTANT
Check that the mains voltage corresponds to the voltage stated
on the type plate on the back of the machine.
Changing the voltage setting
If the factory setting is not the correct setting for your mains supply the
setting can be changed from 240V to 220V: