Error Messages on the Control Panel
Over temp:
If the unit exceeds 195
o
F (91
o
C) turn off main power and contact technical service at
Low water:
1.
If there is water in the reservoir, make sure that the tank is NOT filled exclusively with
distilled water. The unit must have some tap water in it for the sensor to register the water
level. If there is tap water in the reservoir, turn off the main power and contact technical
service. The recommended amount of tap water is 1 cup/gallon of distilled water.
2.
If there is no water in the reservoir, turn off the main power and investigate possible
causes - water evaporation or unit leakage.
•
Water evaporation
- always maintain adequate water volume in the tank.
•
Water leakage
- turn off main power and contact technical service.
UV lamp defective:
Turn main power off. Unplug cord from unit and wall. Remove rear access panel via 4 screws.
Carefully remove black rubber boot from UV housing, allowing the lamp to slide out of the
housing. Disconnect lamp from connectors on both ends. Replace bulb. Reconnect at both ends
and insert new bulb back into the housing. Attach rear access panel. Dispose of defective bulb
according to local regulation.
Temp sensor error:
Refers to the water temperature exceeding 195
o
F (90.5
o
C). If the over temp error message displays:
1.
While evacuating the water, pause evacuation, wait for the water to cool, then resume
evacuation. Unplug cord from unit and wall. Contact technical service.
2.
During standard operation, unplug cord from unit and wall. Contact technical service.
Storage and Handling
•
Check the power cord and plug for excessive wear or fraying. Replace as needed.
Do not use a detachable mains power cord with inadequate ratings. Only replace
with a Strava power cord.
It is recommended the following guidelines are used whenever this system is being stored
or transported to another location:
Operation: 32
o
F – 104
o
F (0
o
C – 40
o
C)
Storage: 32
o
F – 140
o
F (0
o
C – 60
o
C)
2
Pollution degree
Operation: Up to 80%
Storage: Up to 90%, non-condensing
Environment
(Temperature)
Environment (Humidity)