![Storz 40712035 Instructions For Use Manual Download Page 30](http://html1.mh-extra.com/html/storz/40712035/40712035_instructions-for-use-manual_1382978030.webp)
Fault correction
Instructions for use • DRILLCUT-X II/DRILLCUT-X II-35 • WWK497_EN_V1.0_03-2020_IFU_CE-MDR
30
10 Fault correction
10.1 Troubleshooting
Symptom
Possible cause
Remedy
Complete failure of the device Power supply failure.
Check power supply.
Power cord plug is not prop-
erly connected to the socket
on the device.
Push the power supply
plug firmly into the
socket on the device.
Defective power cord.
Replace power cable.
Instruments do not rotate.
Connection to motor inter-
rupted.
Notify customer service.
No power output when
footswitch or keys pressed.
Handpiece is damaged.
Connect another
handpiece.
Handpiece blocked or
jammed.
Clean handpiece.
Speed cannot be regulated.
Device is possibly damaged.
Notify customer service.