ASTOPAD
®
Instructions for Use
38
10.1 Low temperature alarm “A1”
(low priority alarm condition)
Control panel
Device response
•
The temperature reduces in controlled state to 1.0 °C
below the set temperature
.
•
If this condition lasts for more than 10 minutes, the
acoustic alarm is triggered, the “Alarm” LED
turns on
and the display
shows “A1”.
Alarm condition
This alarm is displayed when the difference of the applied
part’s actual temperature
to the set temperature is
≥ 1.0 °C lower.
Possible reasons
►Required
action(s)
The applied part has not yet reached the set temperature.
►Warm-up phase, no action necessary.
Applied part defective.
►Send the applied part to the local sales point.
Required action(s)
for resetting
Press the “Start”
button to return the ASTOPAD to the
control mode. The acoustic alarm signal is then sup-
pressed for a further 10 minutes and the display
shows
the symbol “Applied part heating” . If the alarm condition
recurs, there is a defect in the applied part.
As soon as the set temperature is reached or the alarm is
reset, the acoustic low-temperature alarm “A1” will be sup-
pressed for 10 minutes according to IEC/EN 80601-2-35.
2
10
5
1
2
6
5