7 Technical data
ASTODIA Instructions for Use
23
7 Technical data
DIA120
Control Unit
DIA130
Hand piece
DIA140
Charger
Electrical connection
-
-
100-240VAC ±10%
50-60Hz
Nominal voltage
4,8 V
-
-
Classification
(IEC/EN 60529)
IPX0
IP4X
Protection class
-
-
Isolation class II
Classification
(MDD 93/42/EEC)
Class I
Code UMDNS
14-130
Dimensions
approx. Height
approx. Width
approx. Depth
141 mm
63 mm
33 mm
70 mm
20 mm
10 mm
100 mm
45 mm
75 mm
Weight
0,2 kg
0,03 kg
0,2 kg
Operating mode
Continuous operation
Battery type
Battery pack (4x AA, Ni-MH, 4,8 V, min. 1900 mAh)
Battery capacity
Approx. 5 hours at the highest level
Charging time
Approx. 3 hours
Essential performance
according to IEC/EN
60601-1
Control of the brightness of the LEDs of the hand-
piece in different stages.
Automatically switch
over
If the device is constantly operated with the highest
light level, it will automatically switch over to the next
lower level after 2 minutes.
Colour LED
Red: 620 – 640 nm
Yellow: 584 – 597 nm
Illumination intensity
At least 19 lumens (at the highest level)
Permissible ambient
conditions
Humidity
Temperature
Atmospheric
pressure
In operation
20% to 85%
non-condensing
+10 °C to +40 °C
700 hPa to 1060 hPa
During storage
20% to 90%
non-condensing
-20 °C to +60 °C
500 hPa to 1060 hPa
DANGER
Explosion hazard!
Do not use the ASTODIA diaphanoscope in an explosion-hazard environment
or in the presence of flammable anaesthetics.
Summary of Contents for ASTODIA DIA100