![Steris SYSTEM 1E Maintenance Manual Download Page 93](http://html.mh-extra.com/html/steris/system-1e/system-1e_maintenance-manual_1374323093.webp)
7-16
764333-675
7.20 CHANGING STERILE AIR FILTER
WARNING – PERSONAL INJURY
HAZARD: Disconnect all utilities to
SYSTEM 1E Liquid Chemical Sterilant
Processing System before servicing. Do not
service the SYSTEM 1E Liquid Chemical Sterilant
Processing System unless all utilities have been
properly locked out. Always follow OSHA Lockout-
Tagout and electrical safety-related work practice
standards.
WARNING – PERSONAL INJURY and/or
EQUIPMENT DAMAGE HAZARD: Repairs
and adjustments to this equipment must
be made only by fully qualified service personnel.
Maintenance performed by inexperienced,
unqualified personnel or installation of
unauthorized parts could cause personal injury,
invalidate the warranty, or result in costly
equipment damage. Contact STERIS regarding
service options.
Procedure:
The sterile air filter should be changed every six
months.
The sterile air filter must be installed in the
SYSTEM 1E Processor prior to operation.
1. Open control panel and locate sterile air filter.
2. Remove air filter.
Pull up on metal ring on top of the filter while
twisting to remove filter.
3. Replace with new air filter.
With flow arrow pointing down, turn filter
clockwise.
4. Close control panel.
!
!
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......