4-19
Processing Instructions
Operator Manual
920011-649
4.4.4 Status
See Figure 4-20.
When an automated leak test is performed and leak test hose(s) is
(are) connected, no further action is required. The SINGLE or DOUBLE
cycle runs and the following printout is printed immediately after the
leak test sequence:
========================
HH:MM:SS XM
========================
LEAK TEST 1: PASS
LEAK TEST 2: PASS
========================
or:
========================
HH:MM:SS XM
========================
LEAK TEST 1: PASS
LEAK TEST 2: NONE
========================
If the automated leak test fails, an alarm is triggered (Fault 52 or
Fault 53; refer to
S
ECTION
7, T
ROUBLESHOOTING
) and the cycle is
aborted. A failed leak test indicates there is a leak in the endoscope.
If the automated leak test fails, an alarm sounds and display shows:
or:
alternating with:
and printer message is:
=========================
HH:MM:SS XM
MM-DD-YY
=========================
= FAULT 52
=
=========================
FAULT 52
CYCLE DELAYED
FAULT 53
CYCLE DELAYED
SEE FAULT CHART
CYCLE DELAYED
CAUTION – POSSIBLE EQUIPMENT
DAMAGE:
Remove
any
automated leak test hoses
present
in
endoscope
processing support before
performing a D-SHORT or a D-
LONG cycle. Performing a
decontamination
cycle
(D-
SHORT and D-LONG) with
automated leak test hose
assemblies
present
may
cause water to enter hoses.
Inspect
leak
test
hose
assemblies for presence of
water prior to processing a
cycle. Water in tubing may
cause damage to endoscopes.
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...