3-3
Component Identification
Operator Manual
920011-649
3.1.5 Chamber
See Figure 3-3.
The high-quality stainless-steel chamber is insulated to reduce heat
loss and noise level. Four manifold connectors are provided at the
back of the chamber: one is the hot water inlet to the Reliance
™
CDS
Chemical Delivery System, two are connectors for the pressure
sensors and one is for the endoscope processing support. The
endoscope processing support is automatically connected to these
manifold connectors when positioned correctly in the processor
chamber.
A removable debris screen at the bottom of the chamber prevents
large debris from entering the piping system and pump. The screen
is provided with a handle and is easily removed for cleaning.
An interior fluorescent light allows the operator to view inside the
chamber during a Reliance EPS Endoscope Processing Cycle.
3.1.6 0.2 µm Bacterial-
Retentive Filter
See Figure 3-3.
All the water supplied to the processor is filtered through a 0.2 μm
bacterial-retentive filter which is conveniently positioned behind the
front service access panel, below the chamber door, for easy
maintenance.
3.1.7 Cleaner Injection
Pumps
See Figure 3-3.
The peristaltic cleaner injection pumps automatically inject the correct
quantity of cleaner during the appropriate cycle phase. The pumps
are conveniently positioned behind the front service access panel,
for easy maintenance.
3.1.8 Fresh Container
Detection Meter
See Figure 3-3.
The fresh container detection meter is used to detect the presence
of the Reliance dry chemistry container during SINGLE and DOUBLE
cycles.
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...