![Steris InnoWave PCF Sonic Irrigator Operator'S Manual Download Page 112](http://html.mh-extra.com/html/steris/innowave-pcf-sonic-irrigator/innowave-pcf-sonic-irrigator_operators-manual_1374347112.webp)
9.1
Rewash Program Setting
Within a wash program, if the rewash parameter is set above zero, the rewash
feature is enabled. When the rewash feature is enabled, the tank fluids do not drain
at the end of the first wash cycle. The operator then has the option to reuse the tank
fluids for the next wash cycle.
If the sonic irrigator has editable (open) wash programs, the rewash parameter can
be set and adjusted through the supervisor menu. Some configurations of the sonic
irrigator may ship with one or more factory-set rewash programs. The equipment
owner can use the supervisor menu to view program settings and see what rewash
options are available. See
Viewing and Editing Wash Program Parameters
The rewash feature is intended for heavily soiled instruments only. On completion
of a rewash cycle, it is strongly recommended that instruments are further
processed in a standard wash cycle to rinse off any remaining soil and chemicals.
WARNING
If setting up a rewash program for use with instruments contaminated with
blood, it is recommended that the wash temperature is no higher than +35 ºC
(+95 ºF). High- temperature washes without a pre- wash cycle can leave
clotted blood on instruments and can promote formation of prions, leading to
instrument contamination.
When performing a succession of rewash cycles, repeated sonic activity may
increase the tank water temperature, and care should be taken to not exceed
the recommended limit of +35 ºC (+95 ºF) at any stage.
9.2
The Rewash Counter
When using the rewash option, the display will
show a number after the washing stage name
to indicate how many times the same tank fluids
have been used.
When the number of times the tank fluids have
been used reaches the number set in the
rewash program parameters, the tank drains at
the end of that cycle, and the operator is not
given the option to reuse the tank fluids.
104
THE REWASH FEATURE
InnoWave PCF Sonic Irrigator
Operator Manual
Summary of Contents for InnoWave PCF Sonic Irrigator
Page 1: ...InnoWave PCF Sonic Irrigator Operator Manual ENGLISH ...
Page 2: ......
Page 20: ...12 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 86: ...78 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 98: ...90 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 110: ...102 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 120: ...112 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 138: ...130 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 156: ...148 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Page 160: ...152 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...
Page 162: ...154 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...
Page 164: ...156 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...