9.4.2
Installing Argon Gas Bottle
1. Remove the argon gas bottle from the shipping box. Inspect for shipping damage.
2. Remove the red cap from the argon gas bottle and discard.
3. Turn the gi4000 ESU ON by pushing the power ON/OFF switch to the ON position (the
“I” should be pressed inward).
4. Select the Argon Method on the touchscreen display.
5. Carefully align and insert the argon gas bottle, valve end first, into the port located on
the right-hand side of the gi4000 ESU.
6. Screw in the argon gas bottle
clockwise
until it is fully seated, and you observe a
green gas bottle icon appear above the gas information section on the left side of the
Argon Method screen.
7. Confirm that the argon gas bottle icon on the Argon Method screen displays a full,
green bottle icon.
WARNING: Personal Injury Hazard.
Any attempt to operate the gi4000
ESU with any gas source other than the gi4000 argon gas bottle, or to
attach argon probe accessories to the gi4000 ESU in any way other than
by the use of the ArConnect
®
argon probe connector may pose a serious
hazard to patients and users.
NOTE:
It is recommended that the water bottle holder be removed during
installation and removal of the argon gas bottle for easier access to the
argon port.
NOTE:
Continue turning clockwise firmly until the argon gas bottle is fully
engaged. This can be anywhere from ¼ to 2 full turns tighter, even after
seeing the green gas bottle icon appear. Do not stop turning until the
argon gas bottle is fully seated and cannot be turned any tighter to
prevent argon from leaking from the system at the bottle valve.
NOTE:
If the GREEN gas bottle icon does not appear, remove the
argon gas bottle by turning
counterclockwise
and take it out of the
gi4000 ESU. Inspect the brass threads for damage. If the threads are
undamaged, while using gloved hands, pour a few drops of sterile water
onto clean, lint-free gauze and lightly wipe around the brass threads.
Repeat steps 5 through 7 above.
Summary of Contents for G1110001
Page 1: ...gi4000 Electrosurgical Unit Reference Reorder No G1110001 User and Maintenance Manual ...
Page 6: ... This page is intentionally left blank ...
Page 54: ...00732467 Rev D Page 48 of 110 8 11 9 Bipolar Method Figure 8 25 Figure 8 26 Figure 8 27 ...
Page 109: ...00732467 Rev D Page 103 of 110 Figure 13 21 Argon RF Output Test Diagram ...