3-7
764333-268
• State of MOSFET that drives hydraulic pump
commands (if MOSFET is shorted, motherboard
can detect it and turn OFF power to pump).
4. Master control board sends following signals to
power supply controller:
• Signal to turn bar graph ON and OFF.
• Command MOSFET to conduct power to pump
(driving with PWM signal).
5. 230 V Power Supply Variation:
The 120 V power supply shows the battery
charge level on the bar graph when on battery
power. But, the 230 V power supply only
shows the battery charge level on the bar graph
when the system is operating on facility (ac)
power. When the system is operating on
battery power, the Hand Control icon indicates
the battery charge level in three stages: Full,
Half and Empty.
* Software is NOT downloadable.
†
Requires Cmax 110 Flash Loader to download.
‡
Requires Cmax Service 1.8 or higher to download.
3.5 TECHNICAL DRAWINGS
Located at the end of this manual (behind the tab
Reference Drawings
) are the Cmax Surgical Table
reference drawings. These drawings are supplied as an
aid in understanding unit operation and troubleshooting.
Table 3-1. Cmax 110 Master Control and Hand Control Assembly Compatibility Matrix
Master Control Assembly
Software
Version
Hand Control*
Heavy Rubber
Housing
(P542200-904)
Hard Plastic
Housing
(P150832-500)
P542200-903
†
1.43
Y
Y
P542200-903
†
1.46
Y
Y
P542200-903
†
1.47
Y
Y
P542200-903
†
(NLA)
1.48
Y
Y
P150832-769
‡
2.04
Y
Y
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......