5-5
150830-685
Operator Manual
Routine Maintenance
3.
Wipe caster with a cloth, dampened with water, to remove
cleaning fluid and debris. Ensure items such as suture, oils, and
floor wax are removed.
4.
Perform
Steps 2 and 3
for remaining three casters.
5.5 Battery Charging
Procedure
Cmax Surgical Table is equipped with INTELLIPOWER
®
dual power
system. The table is powered either by facility power or by internal
batteries. If table is to be operated on battery power, note the
following:
IMPORTANT:
Batteries are recharged automatically as long as the
Main Power Switch is set to
ON
(lighted green) and the ac power is
connected to the Cmax Surgical Table. The PLUG icon appears on
Power Panel LED Display (see Figure 5-1) indicating ac power and
diode bar indicates battery condition.
Bar graph description for 120 V tables:
•
When table is plugged into facility ac and Main Power Switch is
ON, diode bar shows battery charge state. This battery charge
state is also indicated by icon on primary hand control (refer to
Figure 4-4).
•
When table is on battery power and primary hand control is
active (on), diode bar shows battery charge state. This battery
charge state is also indicated by icon on primary hand control
(refer to Figure 4-4).
Bar graph description for 220 V tables:
•
When table is plugged into facility ac and Main Power Switch is
ON, diode bar shows battery charge state. This battery charge
state is also indicated by icon on primary hand control (refer to
Figure 4-4).
•
When table is on battery power and primary hand control is
active (on), diode bar is off. Battery charge state is indicated by
icon on primary hand control (refer to Figure 4-4).
Batteries will require recharging on a periodic basis depending on
frequency of table usage. Low or discharged battery conditions are
indicated by HALF FILLED or EMPTY BATTERY icon on the Hand
Control (refer to Figure 4-4 and
4.2.2, H
AND
C
ONTROL
LCD
D
ISPLAY
D
ESCRIPTION
) and some or all green diode bars on Power
Panel LED Display are OFF.
NOTE: Fully charged batteries enable between 100 and 150
procedures (approximately three weeks) without recharging. Note the
charge/discharge icon on the LED Display (see Figure 5-1) of Power
Panel.
Lead acid batteries last longer if not fully discharged. Therefore, to
obtain the longest life and capacity from your Cmax Surgical Table
batteries, always connect ac power cord to table base and plug into
an appropriate ac receptacle as often as possible, and as long as
possible, with the Main Power Switch set to
ON
(lighted green). If this
is not always possible, recharge batteries at the following times:
NOTE: When two green diode bars on Power Panel LED Display are
OFF, batteries are half charged. This level of power enables normal
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD: Repairs
and adjustments to this
equipment must be made only
by STERIS or STERIS-trained
service personnel.
Maintenance performed by
unqualified personnel or
installation of unauthorized
parts could cause personal
injury, result in improper
equipment performance,
invalidate the warranty, or
result in costly damage.
Contact STERIS regarding
service options.
WARNING – TRIPPING
HAZARD: Route the power
cord to the receptacle in a
position so that it will not be
tripped over by personnel in
the area.
Summary of Contents for Cmax
Page 1: ......
Page 4: ......
Page 65: ......
Page 142: ...Page 3 of 14 ...
Page 148: ...Page 9 of 14 ...
Page 149: ...Page 10 of 14 ...
Page 155: ...Page 2 of 2 Figure 1 Second Spacer ...
Page 164: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......