Operator Manual
Installation Verification
An Equipment Drawing, showing all utility and space requirements, was sent to
you after the order for this equipment was received. The clearance space,
specified on Equipment Drawing (P920502-212), is necessary for proper
installation, operation, and maintenance.
Uncrating/Installation Instructions
(P920017-147) were provided with the washer. If any of these documents are
missing or misplaced, contact STERIS giving the serial, equipment, and model
numbers of the unit. Replacement copies will be sent to you promptly.
After installing the washer according to the
Uncrating and Installation Instruc-
tions (P920017-147), complete the following checklist to assure complete and
correct installation, or contact STERIS to schedule a technician to test your
installation and demonstrate proper equipment operation.
❑
Shutoff valves (not provided by STERIS), for maintenance purposes and
capable of being locked in OFF position only, are installed on steam, air, and
water lines and are in compliance with local occupational health and safety
regulations, as well as plumbing codes.
❑
Disconnect switches (not provided by STERIS), for maintenance purposes
and capable of being locked in OFF position only, are installed on electrical
supply lines near the unit and are in compliance with local occupational health
and safety regulations, as well as electric codes. Disconnect switches must
bear the identification of the equipment to which they are connected.
NOTE: If unit is installed next to other equipment, shutoff valves and disconnect
switches should be located so that service can be shut off to one piece of
equipment at a time.
❑
Washer positioned as shown on Equipment Drawing and with required
clearance space and in relation to building supply lines.
❑
Washer and two mechanical cores are level. Use shims or Leveling Legs (if
accessory is present) if necessary.
❑
If Drain Discharge Cooldown System or pH Neutralizing System option is
present, building cold water line supplies water to unit as specified on
Equipment Drawing.
❑
Building hot water line supplies water to unit as specified on Equipment
Drawing.
❑
Building steam line provides steam to unit as specified on Equipment
Drawing.
❑
Building condensate return line is connected to washer as specified on
Equipment Drawing.
❑
Building air line supplies air to unit as specified on Equipment Drawing.
❑
Building waste line is connected to washer as specified on Equipment
Drawing.
❑
Building ventilation system is connected to washer as specified on Equip-
ment Drawing.
❑
Electrical supply for unit is as specified on Equipment Drawing.
❑
Piping connections between washer, main mechanical core, and secondary
mechanical core are as specified in Section 3 of
Uncrating/Installation
Instructions (P920017-147).
2.2 Installation
Checklist