3-6
764333-098
3.14.4 Alkaline Wash Tank Heating Alarm
If after 10 minutes (default value adjustable in service
mode only), alkaline wash tank water temperature is
not reached, control gives an alarm.
3.15 ACID WASH TANK MONITOR
3.15.1 Acid Wash Tank Filling
If acid tank high water level switch (1 in 10) indicates
that acid wash tank is not full, control energizes acid
tank fill valve (5 out 4) for 5 minutes (default value
adjustable in service mode only) or until acid wash
tank is full. Once acid tank high water level switch
detects water (1 in 10) for 3 consecutive seconds, con-
trol de-energizes acid tank fill valve (5 out 4).
3.15.2 Acid Wash Tank Filling Alarm
If after 5 minutes (default value adjustable in service
mode only), acid tank high water level switch does not
detect water (1 in 10), control gives an alarm.
3.15.3 Acid Wash Tank Heating
Once acid wash tank is full, control energizes acid tank
steam valve (5 out 6) to start heating acid wash tank
water up to 140.0
°
F [60.0
°
C] (default value adjustable
in service mode only) for 10 minutes (default value
adjustable in service mode only). Once temperature set
point is reached, control de-energizes acid tank steam
valve (5 out 6).
3.15.4 Acid Wash Tank Heating Alarm
If after 10 minutes (default value adjustable in service
mode only), acid wash tank water temperature is not
reached, control gives an alarm.
3.16 MISCELLANEOUS NOTES
3.16.1 Bypass Button
A BYPASS button is present on Cycle Status display.
A cycle phase may be bypassed only if Temperature
Guaranteed is not selected. Recirculation, soaking and
exhaust phases can be bypassed. A message is printed
when a phase is bypassed.
3.16.2 Extend Button
An EXTEND button is present on Cycle Status display.
Pressing EXTEND button adds time to remaining
phase time. E.g. if 1 minute remains from a 5 minute
recirculation and the EXTEND button is pressed, the
recirculation timer is extended to 6 minutes.
Recirculation, soaking and exhaust phases can be
extended. A message is printed when a phase is
extended.
3.17 MANUAL MODE DESCRIPTION
User or service technician can manually control certain
washer functions by accessing Manual Control mode.
Manual Control mode is accessible from Operating
mode when washer is not in cycle. By pressing
MAN-
UAL MENU
button, the following sequences are avail-
able in this mode:
3.17.1 Fill
By selecting FILL, filling port valve is activated and
sump fills. Display shows FILLING SUMP. Once sump
level reached, display shows SUMP FULL. If a door is
opened or FILL button is pressed again, sequence is
aborted. Filling valve closes automatically when sump
is full.
3.17.2 Pump/DRV
By selecting PUMP/DRV the following sequence
occurs: sump is filled; then, recirculation pump, jet
valve and header system are activated. Display shows
PUMPING/DRIVING.
If a door is opened or PUMP/DRV button is pressed
again, the sequence is aborted.
3.17.3 Drive
By selecting DRIVE header system is activated. Dis-
play shows DRIVING. If DRIVE button is pressed
again, sequence is aborted.
3.17.4 Drain
By selecting Drain, you can drain the following:
Sump
Acid Tank (appears only if option selected)
Alkaline Tank (appears only if option selected)
Cooldown (appears only if option selected)
• Sump
Sump is filled to ensure pump does not run
dry. Recirculation pump is activated and drain
valve opens. Display shows DRAINING
SUMP.
Once drain time has elapsed, display shows
SUMP EMPTY. If a door is opened or SUMP
button is pressed again, sequence is aborted.
If cooldown tank is present, this one is drained
using manual cooldown tank draining
sequence.
Summary of Contents for Basil 4700
Page 1: ...MAINTENANCE MANUAL Basil 4700 Cage and Rack Washer 2008 10 15 P764333 097 ...
Page 3: ...OPERATOR MANUAL Basil 4700 Cage and Rack Washer 2008 08 14 P920509 645 ...
Page 172: ...2 764333 099 Figure 1 Basil 4700 Cage and Rack Washer Inside Washer Inside Inside Washer ...
Page 174: ...4 764333 099 Figure 2 Cabinet Assembly Left Side Control Shown ...
Page 176: ...6 764333 099 Figure 3 Sump and Floor Grating Assembly ...
Page 178: ...8 764333 099 Figure 4 Door Assembly ...
Page 180: ...10 764333 099 Figure 5 Double Door Interlock System ...
Page 182: ...12 764333 099 Figure 6 Interior Light ...
Page 184: ...14 764333 099 Figure 7 Safety Emergency Stop Cable Assembly ...
Page 186: ...16 764333 099 Figure 8 Spray Header Assembly ...
Page 188: ...18 764333 099 Figure 9 Cable and Roller Assembly DETAIL A DETAIL B See Detail A See Detail B ...
Page 192: ...22 764333 099 Figure 11 Automatic Water Rack Flush System For Two Racks Option ...
Page 194: ...24 764333 099 Figure 12 Damper 1 ...
Page 196: ...26 764333 099 Figure 13 Exhaust Fan ...
Page 198: ...28 764333 099 Figure 14 Hot Water Line Components 6 5 4 10 10 7 8 9 7 ...
Page 200: ...30 764333 099 Figure 15 Cold Water Line Components ...
Page 202: ...32 764333 099 Figure 16 Recirculation Piping ...
Page 204: ...34 764333 099 Figure 17 Steam Line Components ...
Page 206: ...36 764333 099 Figure 18 Final Rinse System Assembly ...
Page 210: ...40 764333 099 Figure 20 Pneumatic Block ...
Page 212: ...42 764333 099 Figure 21 Peristaltic Chemical Injection Pumps ...
Page 214: ...44 764333 099 Figure 22 Solution Tank 25 ...
Page 216: ...46 764333 099 Figure 23 Tank Support Frame ...
Page 218: ...48 764333 099 Figure 24 Guaranteed Cooldown Tank ...
Page 220: ...50 764333 099 Figure 25a Electrical Box and Control Column Assembly ...
Page 226: ...56 764333 099 Figure 26 Control Panel See Detail A DETAIL A ...
Page 228: ...58 764333 099 Figure 27 Seismic Tie Down Anchorage Option ...
Page 230: ...60 764333 099 Figure 28 Bottle Washing Cart WA02 010 ...
Page 232: ...62 764333 099 Figure 29 Bottle Washing Cart WA02 020 ...
Page 234: ...64 764333 099 Figure 30 Pan Cart WA02 030 Accessory ...
Page 236: ...66 764333 099 Figure 31 Universal Cage and Pan Wash Cart WA03 100 7 ...
Page 238: ...68 764333 099 Figure 32 Universal Cage Wash Cart WA02 040 ...
Page 240: ...70 764333 099 Figure 33 Cart Tilt Ramp WA02 050 ...
Page 242: ...72 764333 099 Figure 34 Floor Ramp for Floor Mounted Units WA02 060 ...
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ......