![Steris AMSCO 400 Series Operator'S Manual Download Page 82](http://html.mh-extra.com/html/steris/amsco-400-series/amsco-400-series_operators-manual_1374315082.webp)
6-22
P129394-020
Operator Manual
Sterilizer Operation
Table 6-4. Factory-Set
SFPP
Sterilizer Cycles and Cycle Values
Cycles
Sterilize
Temp.
Sterilize
Time
Dry
Time
Recommended Load
Validation
Standard
SFPP Cycle
270°F
(132°C)
4
Minutes
30
Minutes
Double-wrapped instrument
trays, porous and non-
porous loads. Maximum
weight per tray, 25 lb
(11.3 kg)
ST-8:2008
SFPP Cycle
270°F
(132°C)
4
Minutes
20
Minutes
*
Fabric Packs
ST-8:2008
SFPP Cycle
275°F
(135°C)
3
Minutes
30
Minutes
Double-wrapped instrument
trays, porous and
non-porous loads.
Maximum weight per tray,
25 lb (11.3 kg)
ST-8:2008
*
Dry time can be reduced to 5 minutes for a single fabric pack.
Figure 6-11. SFPP Cycle In-Cycle Screens
Cycle Selection Screen
Cycle Selection Screen
Condition Phase
Sterilize Phase
Exhaust Phase
NOTE: Countdown time on display is estimated; timer self-corrects estimated time at beginning
of each phase.
Press for More Cycles
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...