![Steris AMSCO 400 Series Operator'S Manual Download Page 75](http://html.mh-extra.com/html/steris/amsco-400-series/amsco-400-series_operators-manual_1374315075.webp)
6-15
Sterilizer Operation
Operator Manual
P129394-020
6.10 Gravity Cycles
The Gravity cycle is used for sterilizing one or more unwrapped
instrument trays containing non-porous items.
1.
Refer to
2.
See instructions for using the loading car/transfer carriage or
rack and shelves, earlier in this section, for procedures to load
the sterilizer.
3.
To start the Gravity cycle, press the appropriate
GRAVITY
touch-screen pad. Refer to
for descriptions of cycle
use.
4.
Confirmation screen #47 or #51 appears on the control (see
facing page). Verify correct load is in the chamber to match
sterilize exposure time for selected cycle and press YES or NO
to continue.
NOTE: If the wrong cycle has been selected, see
5.
Sterilizer automatically progresses through cycle, as follows:
NOTE: The control display indicates cycle progress in two ways:
a.
Large numerals in the screen’s center display estimated
remaining cycle time,
b.
Change of phase name in the top status line of the screen.
ACTIVATE SEAL – Steam enters the door seal, pressing it
against inside surface of the door.
PURGE – Chamber is purged with steam. Start of condition is
printed.
CHARGE – Chamber is charged with steam. Start of steam
charge is printed.
STERILIZE – Start of sterilize exposure is printed when the
chamber reaches sterilization temperature. Chamber
temperature is printed every minute.
FAST EXHAUST – Start of exhaust is printed and chamber is
exhausted to 4 psig.
DRY – Start of dry is printed and display counts down dry time
remaining.
AIR BREAK – Chamber is returned to atmospheric pressure.
RETRACT SEAL – A vacuum is drawn on the seal, retracting it
from inner surface of door.
COMPLETE – Complete tone sounds. Cycle summary and end of
cycle messages are printed.
6.
Unload sterilizer (see instructions for using the loading car/
transfer carriage or rack and shelves, earlier in this section for
procedures to remove load).
WARNING – BURN HAZARD:
•
Sterilizer, rack/shelves and
loading equipment will be hot
after cycle is run. Always
wear protective gloves and
apron when removing a
processed load. Protective
gloves and apron must be
worn when reloading
sterilizer following the
previous operation.
•
Steam may be released from
the chamber when door is
opened. Step back from the
sterilizer each time the door
is opened to minimize
contact with steam vapor.
WARNING – SLIPPING
HAZARD: To reduce slippery
conditions, keep floors dry by
immediately wiping up any
spilled liquids or
condensation in sterilizer
loading or unloading area.
WARNING – PERSONAL
INJURY HAZARD: When
closing the chamber door,
keep hands and arms out of
the door opening and make
sure opening is clear of
obstructions.
!
!
!
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...