5-22
P129394-020
Operator Manual
Component Identification
additional time is required to depressurize generator, because
steam cannot exhaust through the door seal. If “Too Long to
Depressurize Generator” alarm occurs, increase the factory
default setting for too long to depressurize.
1
•
Cold water utilities run during the depressurize phase. Minimize
utilities usage by closing the operating end door.
2.
Drain
– Automated drain ball valve opens, allowing hot water
and steam to flow from generator to drain. Cooling water flows
continuously to the generator drain to maintain an average of
140°F (60°C) or less. The generator drain valve must open within
one minute or an alarm occurs. A 10-second timer appears on
the status bar when conditions for Drain phase have been met.
3.
Flush
– Hot water flows through boiler chamber for 5, 10 or 15
minutes as selected by user. Cooling water is controlled by
temperature probe to maintain an average generator drain
temperature of 140°F (60°C) or less.
1.
Alarm intervals (such as “Too Long to Depressurize Generator”) must be
adjusted by trained technicians, only. Contact STERIS for this service.
STATUS . .DRAIN 0:10
4
TEMP . . .168.3 F
PRESS . . .0.0 psig
GENERATOR FLUSH
PROJECTED CYCLE COMPLETION TIME:
5:14
MINUTES
SECONDS
PAPER
STATUS
FEED
ABORT
Figure 5-17. Integral Generator Flush Cycle
Jacket Temperature
Generator Pressure
212°F
Idle
De
pres
surize
Dr
ai
n
Flush
Comple
te
0 psig
Indicates Cycle
Transition
Points Printed
During Cycle
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...