Operator Manual
Component Identification
5.6.2 Abort Touch-Screen
Pad
The Abort touch-screen pad is used to end a cycle before it finishes
normally. A cycle only needs to be aborted if an abnormal condition
or a control problem develops during the cycle. Pressing Abort
causes the sterilizer chamber to depressurize (if pressurized), or Air
Break (if in vacuum); the door seal deactivates, the control prompts
the operator to open the door and the sterilizer returns to its normal
out-of-cycle state. If an abnormal condition persists after fully
aborting the cycle, contact your supervisor or a qualified service
technician before trying to operate the sterilizer further.
Figure 5-9. Cycle Abort Touch-Screen Pad
STATUS . . DRY 0:45
4
TEMP . . . 210.0 F
PRESS . . . 0.29 psig
3.TRAY, PREVAC, 270F, 4, 30
PROJECTED CYCLE COMPLETION TIME:
03:18
MINUTES
SECONDS
PAPER
STATUS
FEED
ABORT
Press this
button to abort
a cycle
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...