1-1
Safety Precautions
Operator Manual
P129394-020
1
The following
Safety Precautions
must be observed when operating or servicing this AMSCO
®
400 Series Small
Steam Sterilizer. WARNING indicates the potential for personal injury and CAUTION indicates the potential for
damage to equipment. For emphasis, certain
Safety Precautions
are repeated throughout the manual. It is
important to review
all
Safety Precautions
before operating or servicing the unit.
Strictly following these
Safety Precautions
enhances the Customer’s ability to safely and effectively utilize the unit
and helps avoid improper maintenance methods which may damage the unit or render it unsafe. It is important to
understand that these
Safety Precautions
are not exhaustive; Customers are encouraged to develop their own
safety policies and procedures to enhance and complement these
Safety Precautions.
Important:
Please read this Equipment Manual to get the best understanding and to help ensure best
performance of this equipment, which has been throughly inspected, tested and adjusted by factory-trained
personnel.
• Equipment maintenance other than described in the manual (unless by authorized representatives of the
STERIS Corporation), shall be at the risk of invalidating the equipment warranty.
• To obtain authorized service call:
STERIS Corporation
Healthcare: 1-800-548-4873
Life Sciences: 1-800-444-9009
WARNING – ELECTRIC SHOCK AND BURN HAZARD:
WARNING – PERSONAL INJURY HAZARD:
WARNING:
Sterilizers may be equipped with up to three electrical connection points (sterilizer control,
vacuum pump and/or integral steam generator). Each electrical supply must be provided with a
separate disconnect. Disconnect all utilities to sterilizer before servicing.
Do not service the
sterilizer unless all utilities have been properly locked out. Always follow OSHA Lockout-Tagout and
electrical safety-related work practice standards.
Avoid personal injury from bursting bottles.
Liquid sterilization cycle must only be used for liquids in
borosilicate (Pyrex
®
a
) flasks with vented closures.
a. PYREX
®
is a registered trademark of Corning Incorporated.
The liquid cycle is for non-patient contact use only.
!
!
!
SAFETY PRECAUTIONS
1
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...