9-31
Troubleshooting
Operator Manual
P129394-020
Alarm
Description
Screen with Operator Instructions
9.5.12 Generator
Pressure Switch
Failure
Occurs if generator
drain pressure switch is
closed when jacket is
charged.
Screen with Service Instructions
STATUS . . ALARM!
249
GENERATOR PRESSURE SWITCH FAILURE
CHAMBER: 104.3 F 5.4 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
250
GENERATOR PRESSURE SWITCH FAILURE
-> GENERATOR PRESSURE SWITCH
IS CLOSED WHEN JACKET IS CHARGED
CAUSES AND CORRECTIONS:
1. GENERATOR PRESSURE SWITCH FAILURE
-> READJUST GPS SWITCH
-> REPAIR GPS SWITCH
-> REPAIR WIRING
2. GENERATOR PROBE FAILURE
-> REPLACE
-> RECALIBRATE
3. GENERATOR HEATER FAILURE
-> REPAIR GHT
PAPER
FEED
EXIT
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...