7-1
Troubleshooting
Operator Manual
150830-353
1. Cannot turn table ON.
7
TROUBLESHOOTING
This section describes the types of Amsco
®
3085 SP Surgical Table malfunc-
tions most likely to occur, and probable causes and corrective actions. Use the
Operator Troubleshooting Chart to identify general problems. Use the Hand
Control Diagnostics Chart to identify problems as indicated by the hand control
LEDs.
NOTE: Never permit unqualified persons to service the table.
If you are unable to correct the problem with the use of the Operator
Troubleshooting Chart or the Hand Control Diagnostics Chart, or if a problem
occurs not described on the charts, please contact STERIS. A trained service
technician will promptly place your equipment in proper working order.
NOTE: When quickly articulating the table through various repeated movements,
the synchronization of the tabletop sections can get out of alignment and the
hand control may stop functioning properly. To prevent this condition, the user
should perform a return-to-level function after
each
procedure. This will keep the
variance over multiple articulations to a minimum.
PROBLEM
POSSIBLE CAUSE AND CORRECTIVE ACTION
1. Hand control not connected – Connect per
S
ECTION
3
.
2. Hand control defective – Replace.
1. Table unplugged (electric-powered table only) – Plug in.
2. No facility power (electric-powered table only) – Turn facility
power on.
3. F1 or F2 fuse blown (electric-powered table only) – Replace
fuse(s) per
S
ECTION
8
.
4. AC power cord defective (electric-powered table only) –
Replace power cord.
5. Batteries totally discharged (battery-powered table only) –
Recharge batteries per
S
ECTION
6
.
6. Circuit breaker CB-1 tripped (electric-powered table only) –
Reset per
S
ECTION
8
.
7. Circuit breaker CB-2 tripped (battery-powered table only) –
Reset per
S
ECTION
8
.
1. Circuit breaker CB-4 tripped – Reset per
S
ECTION
8
.
2. Circuit breaker CB-2 tripped – Reset per
S
ECTION
8
.
3. AC power cord defective – Replace power cord.
2. No power to pump motor; table will not
articulate.
3. Motor batteries will not charge (battery-
powered table only).
Operator Troubleshooting Chart
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD: Safe and
reliable operation of this
equipment requires regularly
scheduled preventive main-
tenance, in addition to the
faithful performance of rou-
tine maintenance. Contact
STERIS to schedule preven-
tive maintenance.
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......