Document No.: 985-007011
Revision: A
Effective Date: 08/07/2012
Odyssey
Quad HD System Installation
Verification and Testing
Page 18 of 19
“WARNING: This is a revision controlled document. Verify revision is correct prior to use.”
Stereotaxis, Inc
Confidential
4. Peripheral Functionality
4.1.
Required Equipment
•
Laptop computer with external VGA port and spare USB port
•
VGA cable (072-000903-3XX (male to male) or equivalent)
•
USB Cable (072-004656-00XX or equivalent)
•
VGA-DVI Adapter (430-006978-1 or equivalent)
4.2.
Procedure
•
Set the laptop VGA output resolution to 1280 x 1024
•
Enable video output on the laptop VGA connector
•
Verify laptop is set for multiple monitor output
•
Disconnect all cables.
SystemID
Default XML Name
MDM
Channel
USB
Number
USB IP Address
MAC
Addr
TPI S/N
Test Result
1
Live Flouro A
3
1
192.168.168.30
19
Pass
Fail
N.A.
2
Live Flouro B (CR9)
19
2
192.168.168.31
22
Pass
Fail
N.A.
3
Live ECG (CR1)
21
3
192.168.168.32
18
Pass
Fail
N.A.
4
Review ECG (CR2)
12
3
192.168.168.33
15
Pass
Fail
N.A.
5
Carto (CR3)
1
5
192.168.168.34
1C
Pass
Fail
N.A.
6
Ultrasound (PR1)
4
6
192.168.168.35
13
Pass
Fail
N.A.
7
Ref Flouro A
24
7
192.168.168.36
16
Pass
Fail
N.A.
8
Ref Flouro B (CR10)
16
8
192.168.168.37
1A
Pass
Fail
N.A.
9
PR2
14
9
192.168.168.38
10
Pass
Fail
N.A.
10
Leonardo (CR6)
22
10
192.168.168.39
1D
Pass
Fail
N.A.
11
Niobe / CR11
27 / 25
11
192.168.168.40
1E
Pass
Fail
N.A.
13
Video 1 (CR7)
5
12
192.168.168.41
14
Pass
Fail
N.A.
14
Video 2 (CR8)
6
13
192.168.168.42
1F
Pass
Fail
N.A.
15
Video 3 (CR5)
17
14
192.168.168.43
12
Pass
Fail
N.A.
16
Video 4 (CR4)
15
15
192.168.168.44
17
Pass
Fail
N.A.
Table 3 - Connection Cross Reference