1 General information
©
F. Stephan GmbH
GA-602-0106V1.0-PAE-GB
9
1.7 Technical
data
Operation
Temperature
15 – 40 °C
Rel. humidity
30 – 85 %
Air pressure
900 – 1060 hPa
Bring to room temperature before starting to
operate the device.
Storage
Temperature
5 – 60 °C
Rel. humidity
10 – 100 %
Air pressure
700 – 1060 hPa
Store in a protected, dust-free place protected
from moisture and frost
MPG class
Protection class
Inspection/maintenance
cycle
Dimensions (WxHxD)
Weight
II b
I type B as per DIN EN 60601-1 :March 1996
annual
610 x 820 x 680 mm
122 kg
Mains
Connection
230 V AC
50
–
60
Hz
Input power
6 A
Device fuses
2 x 8 AT
2 x 1.25 AT
Absorber unit
Pressure
2.2 bar
Changeover cycle
11 s
Flushing time
2 s
O
2
AIR
System pressure
5.2 bar
5.2 bar
Output pressure
3.8 bar
3.8 bar
Consumption
6 l/min
8 l/min
Max. consumption
7 l/min
13 l/min
Ambient conditions
General
Power supply
Specifications