4 Preparing for operation
20 GA-602-0106V1.0-PAE-GB
©
F. Stephan GmbH
4.2
Connecting to the gas and power supply
4.2.1 Gas
supply
Check the pressure gauge on the front to see whether there is sufficient
pressure. The needle must show more than 3 bar.
The outlets for O
2
and AIR are on the two sides of
S
TAXEL
3.5
. The
connections are quick-release couplings which lock automatically when
connected.
To pull the tube out of the connection, at the same time press the
unlocking pin on top of the connection.
The outlet for vacuum is on the right of the device. The connection is a
1/8" tube nozzle. Simply insert the tube to the intake unit.
4.2.2 Power
supply
S
TAXEL
3.5
runs on 230 V AC.
1. Connect the mains lead at the rear of the device to the power socket
on the wall.
2. Switch on the power switch "On/Off" for O
2
and AIR on the front of
the device. The switch lights up green.
3. Switch the power switch "On/Off" for VAC on the front of the device.
The switch lights up green.
Electrical accessories can be connected to the 4 auxiliary power sockets.
The auxiliary power sockets are switched on with the mains switch
"On/Off" for O
2
and AIR on the front of the device.
When in use,
S
TAXEL
3.5
can only have max. 4 user-accessible auxiliary
power sockets.
Connections for O
2
and
AIR
Connection for VAC
Electrical accessories
Warning