7 Ventilation modes
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
165
NOTE
A two-curve view is always active in the CPAP ventilation mode. The pressure curve
as well as optionally the curve for flow, CO
2
or Pleth is displayed.
Ventilation mode Adjustable parameters
CPAP with PSV
O
2
PEEP
Positive end expiratory pressure (PEEP)
Pressure support (
Psupp)
(relative set value with reference to PEEP)
Expiratory trigger sensitivity (ETS)
Tab. 39: CPAP with PSV
nCPAP
7.3.3.2
The ventilation form nCPAP corresponds to the behavior of the CPAP.
It is only available in non-invasive Neo mode. The device is operated
without an external flow sensor (PNT) in this mode. The pressure is
measured proximally via the pressure measurement adapter set.
Ventilation mode Adjustable parameters
nCPAP
O
2
PEEP
Trigger
Backup
Flow limitation (in NEO mode only)
Tab. 40: nCPAP