background image

STEINDORFF

®

 

 

Fig. 12 

 

 

 

3.1.

 

Adjusting Binocular Viewing Head 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3.

 

ADJUSTMENT AND OPERATION 

Fig. 10 

 

 

 

 

 

Fig. 11 
 

3.1.1.

 

Adjusting interpupillary distance (Fig. 10) 

Different users have different interpupillary distances. So when users 

change, do the interpupillary distance adjustment. 

 

3.1.2.

 

Adjusting diopter ring (Fig. 11, Fig. 12) 

1.

 

Set  the  diopter  rings  of  both  eyepieces  to  “0”position. 

(Do this when users change, because different users have 

different diopters.) 

2.

 

Place an easy-to-observe specimen on the stage plate. 

3.

 

Rotate  the  zoom  body  knob 

  to  the  highest 

magnification 5× (or 6.4×, or 10×), and turn the focusing 

knob 

 to focus the specimen.  

4.

 

Rotate  the  zoom  body  knob 

  to  the  lowest 

magnification 0.8×, looking only into the left eyepiece, 

adjust  the  diopter  ring  on  left  eyepiece  to  focus  the 

specimen. Then do the same thing for the right one. 

5.

 

Repeat item 3 and 4 until the specimen image can always 

focus  without  any  adverse  effect  of  the  changes  of 

magnification on the definition of the image. 

Fig. 11 shows the focusing part of Stereo Microscope 

NYMCS-360/361; 

Fig.12 shows the focusing part of Stereo Microscope 

NYMCS-360. There are coarse focusing knob 

 and 

fine focusing knob 

 in this kind of microscope. 

 

The working distance of the objectives listed in standard 

outfit  is  all  78mm  for  these  three  kinds  of  stereo 

microscopes. 

When 2× objective is chosen, it is necessary to mount the 

binocular  head  bracket 

  onto  the  lower  position. 

Remove  the  clamping  bolt  with  provided  hexagonal 

driver and move the bracket down to the lower position 

and set it with original bolt (Fig. 11). 

Summary of Contents for NYMCS-360

Page 1: ...NYMCS 360 To ensure the safety obtain optimum performance and to familiarize yourself fully with the use of this microscope it is recommended strongly that you study this manual thoroughly before usi...

Page 2: ...justment and Operation 8 3 1 Adjusting Binocular Viewing Head 8 3 2 Focus Adjustment 9 4 Photography and Video Attachment 10 4 1 Photography and Video Attachment Assembly 10 4 2 The Magnification of P...

Page 3: ...t is forbidden to disassemble them yourself 2 2 The binocular viewing head and focusing unit have a compact and precise frame please don t disassemble them as possibly as you can 2 3 Keep the instrume...

Page 4: ...ereo Microscope NYMCS 360 1 COMPONENTS Adjustable WF10 eyepieces Focusing Assembly Binocular viewing head Zoom body Clamping Knob Objectives Glass stage Standard base Brightness adjusting knob Power c...

Page 5: ...ndicate the order of assembly When assembling the microscope make sure that all parts are free of dust and dirt and avoid scratching any parts or touching glass surfaces 2 ASSEMBLY Power cord Microsco...

Page 6: ...om of the viewing head into the mounting port on the zoom body and tighten the viewing head clamping screw with provided hexagonal driver Fig 1 Fig 2 Positioning Pin Mounting Port Note Working Environ...

Page 7: ...sses from being scratched by the eyepiece as in Fig 5 When Not Wearing Eyeglasses Extend the folded eyeshades This makes observation easier by preventing the inverse incidence of light from between th...

Page 8: ...in faintness in field of view In this case adjust the aperture iris diaphragm to the half open state Closing the aperture iris diaphragm to the minimum state and increase the magnification of the zoom...

Page 9: ...8 looking only into the left eyepiece adjust the diopter ring on left eyepiece to focus the specimen Then do the same thing for the right one 5 Repeat item 3 and 4 until the specimen image can always...

Page 10: ...justment knob is used to adjust the top illumination 3 2 2 Adjusting tension collar Fig 15 1 For coarse and fine focus stand if the microscope body falls down by its own weight or the focus obtained b...

Page 11: ...and remove the binocular head away from the zoom body 2 Mount the light splitter unit onto the zoom body and fix it with bolt 3 Mount binocular head onto the light splitter unit with bolt as shown in...

Page 12: ...ication of adapter 1 5 4 3 Switching the Light Path Pull out the light switching lever to open the light path as in figure 19 Light flux ratio Binocular photography 8 2 Zoom Ratio 1 8 Objective Magnif...

Page 13: ...Complete diopter adjustment Dirt is visible in field of view Dirt on specimen Clean specimen Dirt on eyepiece Clean eyepiece The visibility of the image is poor Dust on objective front lens Clean len...

Page 14: ...ED 1 Reflected illuminator LED 1 Annular illuminator LED Fluorescence Eyepiece 10 Eyepiece 2 15 Eyepiece 20 Eyepiece 25 Eyepiece Objectives 0 5 objective 1 objective 1 2 objective Aperture iris diaphr...

Reviews: