7.4.1.
E : 1
– Sensor Zero
Refer to section 8.1.1. for sensor calibration details.
7.4.2.
E : 2
– Sensor Span
Refer to section 8.1.2. for sensor calibration details.
7.4.3.
E : 3
– Sensor Fsd
This Feature sets the FSD of the instrument. It must be matched to the
sensor.
•
Press MENU to open the menu system.
•
Using the
or
buttons, select menu option:
E:3
•
Press
.
•
Using the
or
buttons, change the display to the required
setting.
•
Press
to store the new value.
Note: Pressing the MENU button rather than the
button
exits without any change.
•
Press MENU to close the menu system.
7.4.4.
E : 4
– Output Zero (4mA)
Refer to section 8.2.1. for output calibration details.
7.4.5.
E : 5
– Output Span (20mA)
Refer to section 8.2.2. for output calibration details.