Installation, Commissioning & Routine Gas Testing
FGD10A Oxygen / Toxic Gas Detector
TD18/002
Issue:
16
Change Note:
1904
Page 41
7.5.1. Read example - read live data
The following example is based on V4.x.xE where the data is either byte or
integer types only.
Send the following bytes:
DLE, RD, Variable ID, DLE, EOF, Checksum High byte, Checksum low
byte i.e.
0x10, 0x13, 0x01, 0x10, 0x1F, 0x00, 0x53
Device response on success:
DLE, DAT, Data length
, Data,
DLE, EOF, Checksum High byte, Checksum
low byte
, i.e.
0x10
DLE
0x1A
DAT
0x0D
Data length
0x03
Version
0x01
Display mode
0x07
Sensor Type
0x00, 0x00
Gas reading (signed integer x 100) = 0
0xAE, 0x0B
Current sensor temperature (signed integer x 100) =
29.9
0x92, 0x08
Active sensor signal (unsigned integer) = 2194
0xD7, 0x04
Reference sensor signal (unsigned integer) = 1239
0x10
DLE
0x1F
EOF
0x02
Checksum high byte
0x9F
Checksum low byte