STATUS SCIENTIFIC CONTROLS
Installation, Commissioning & Routine Gas Testing
FGD10A Oxygen / Toxic Gas Detector
TD18/002
Issue:
16
Change Note:
1904
Page 25
4.8.8.
E : 8
– Sensor Selection
This menu option allows the sensor type to be manually selected. The
available sensor options are listed below:
OPTION
TYPE
RANGE
COMMENTS
O2
Electro
chemical
0-25 %v/v
Oxygen
ELT1
Electro
chemical
0-5 ppm
Toxic
ELT2
Electro
chemical
0-10 ppm
Toxic
ELT3
Electro
chemical
0-20 ppm
Toxic
ELT4
Electro
chemical
0-50ppm
Toxic
ELT5
Electro
chemical
0-200ppm
Toxic
ELT6
Electro
chemical
0-500ppm
Toxic
ELT7
Electro
chemical
0-1000ppm
Toxic
ELT8
Electro
chemical
0-9999ppm
Toxic
The sensor type is selected as follows:
•
From the menu system select menu option:
E: 8
and press
ENTER.
•
Using the NEXT and PREVIOUS buttons, select the required
sensor.
•
Press ENTER
Note: Pressing ENTER whilst in menu option
E: 8
will reset the
factory defaults for the displayed sensor. A calibration must be
performed even if the sensor type was not changed.
•
Press MENU to close the menu system.
Note: It is important that the sensor selected via this menu option
corresponds to the sensor that is installed. Incorrect settings here may result
in the inability of the instrument to detect gas.