STATUS SCIENTIFIC CONTROLS
Installation, Commissioning & Routine Gas Testing
FGD10A Oxygen / Toxic Gas Detector
TD18/002
Issue:
16
Change Note:
1904
Page 24
4.8.6.
E : 6
– Decimal Places
The number of decimal places that are used to display the detected gas
levels can be changed via this menu option.
•
From the menu system select menu option:
E: 6
and press
ENTER.
•
Using the NEXT and PREVIOUS buttons, select the required
number of decimal places.
•
Press ENTER.
Note: Pressing MENU instead of ENTER will exit the menu
without changing the number of decimal places to which the
gas level will be displayed.
•
Press MENU to close the menu system.
Note that when the sensor type is first selected, the optimum number of
decimal places for a given sensor type and range is automatically selected.
Increasing the number of decimal places to which detected gas levels are
displayed does not increase the sensitivity or accuracy of the instrument and
in some cases may degrade the instruments perceived stability.
4.8.7.
E : 7
– Firmware Version
The Firmware version is displayed in the form X.Xyy, where X is numerical
and y is a letter.
•
Press MENU to close the menu system.