5.4 Sensor Replacement
Sensor replacement should only be performed by Status Scientific Controls field
service personnel, or a trained engineer. Consult Status Scientific Controls for further
details.
Following sensor replacement:-
•
The sensors electronics must be set-up.
•
A full calibration must be performed, (refer to section 5)
5.4.1
Sensor Electronics Verification.
Select menu mode
C:9
(refer to section 5.2.9)
Select engineer code C:9 - 3 and verify that the sensor temperature is approximately
7
o
C above ambient, @ 25
o
C.
Select engineer code C:9 - 4 and verify that the detector output corresponds with the
table below.
Select engineer code C:9 - 5 and verify that the reference output corresponds with the
table below.
SENSOR
MIN
TYP
MAX
HC
Ref.
TBA
400
TBA
Det.
TBA
700
TBA
Carbon Dioxide
Ref.
750
800
850
Det.
550
600
650
Note:
1) The sensor must be powered for at least 10 minutes before any readings are
taken.
2) The sensors must be in zero gas.