5.2.14
C: 25
– Positive Zero Temperature Compensation
This option is used to apply temperature compensation to the sensor, a zero value
results in no applied compensation.
The value can be either positive or negative, depending upon the type of sensor fitted.
The value is best selected by first calibrating the sensor at 20
C then raising the FGD3
to 50
C. When the temperature is stable, usually after 2 hours, the value is set such
that an accurate zero reading is displayed.
Note:
changing this value from the factory setting may make the readings
inaccurate
– consult Status Scientific Controls Ltd before making any
changes.
•
Press MENU to open the menu system.
•
Using the NEXT and PREVIOUS buttons, select menu option:
C:25
•
Press ENTER.
•
Using the INCREASE and DECREASE buttons, set the required
temperature compensation value.
•
Press ENTER to store the new value.
Note: Pressing the MENU button rather than the ENTER button exits
without any change.
•
Press MENU to close the menu system.
5.2.15 C: 30 - Set Password
Use this menu option to enable / disable the password feature.
Place the FGD9 in the password menu as follows:
•
Press the MENU to open the menu system.
•
Using the NEXT and PREVIOUS buttons, select menu option:
C:30
Note: if the password system is already enabled the user must enter
the 6-digit password to get to this menu.
•
Press ENTER.
•
The display shows either
On
or
OFF
•
Press the UP button to select the desired setting.
•
Press ENTER to accept the setting.
Note:
Pressing MENU instead of ENTER leaves the unit without change.
•
Press MENU to return the instrument to its standard mode of operation.
Note:
if the password is in operation then the user will be prompted with PASS
when ever the menu key is pressed. Pressing the MENU key again will result
in the restricted user access, i.e. only the zero and span options will be
available. Entering the correct password will give access to the full menu
facility.