8.2.68
SensorTemperatureRegulationEffort
Name
SensorTemperatureRegulationEffort
Category
Interface
IInteger
Access
Unit
%
Visibility
Beginner
Values
≥
-100,
≤
100
Indicates the proportion of the available cooling / heating capability
currently being applied to provide stabilization of the sensor tem-
perature. Positive values indicate cooling, negative values indicate
heating.
If the system can not bring the sensor temperature into the selected
range, this value can be checked to identify whether the system
is too hot (maximum cooling effort will be reached) or too cool
(maximum heating effort will be reached) for the chosen
8.2.69
SensorTemperatureRegulationStatus
Name
SensorTemperatureRegulationStatus
Category
Interface
IEnumeration
Access
Unit
-
Visibility
Beginner
Values
InRange, OutOfRange
Indicates state of sensor temperature regulation.
•
InRange
. Regulation is in stable range.
•
OutOfRange
. Regulation is not in the stable range. This can
arise due to a change in
, or a change in environmental conditions, as well as
on power-up.
Usually the status will change to InRange after a period of
time. if it does not, it may be that the selected
is not suitable for the ambi-
70 of 76
Summary of Contents for Swift-G OI-G1480
Page 19: ...Figure 4 1 Overview of StarStream GUI program 11 of 76...
Page 33: ...Figure 6 2 StarStream GUI Property Browser showing valid Persistent IP configuration 25 of 76...
Page 36: ...Figure 6 3 StarStream GUI Connect and Set Static IP dialogs 28 of 76...
Page 44: ...Figure 7 1 Default Trigger Control Figure 7 2 Example Hardware Trigger Control 36 of 76...