Chapter 4
Programming
113
SR860 DSP Lock-in Amplifier
OFLT(?) { i }
The
OFLT i
command sets the time constant according to the table below.
i
time constant
i
time constant
i
time constant
0
1 μs
8 10 ms
16 100 s
1
3 μs
9 30 ms
17 300 s
2
10 μs
10 100 ms
18 1 ks
3
30 μs
11 300 ms
19 3 ks
4
100 μs
12 1 s
20 10 ks
5
300 μs
13 3 s
21 30 ks
6 1 ms
14 10 s
7 3 ms
15 30 s
Example
OFLT 10
Set the time constant to 100 ms
OFLT?
Returns the time constant i
OFSL(?) { i }
The
OFSL i
command sets the filter slope to 6 dB/oct (i=0), 12 dB/oct (i=1), 18 dB/oct
(i=2) or 24 dB/oct (i=3).
Example
OFSL 1
Set the filter slope to 12 dB/oct
OFSL?
Returns the filter slope i
SYNC(?) { OFF | ON | i }
The
SYNC i
command turns the synchronous filter off (i=0) or on (i=1).
Example
SYNC ON
Set the synchronous filter to on
SYNC 1
SYNC?
Returns the state of the synchronous filter i
ADVFILT(?) { OFF | ON | i }
The
ADVFILT i
command turns the advanced filter off (i=0) or on (i=1).
Example
ADVFILT ON
Set the advanced filter to on
ADVFILT 1
ADVFILT?
Returns the state of the advanced filter i
ENBW?
The
ENBW?
query returns the equivalent noise bandwidth of the output filter, in hertz.
Note that the effect of the
SYNC
filter is neglected by
ENBW?
Example
ENBW?
Returns the equivalent noise bandwidth, in hertz.
Summary of Contents for SR860
Page 1: ...Revision 2 01 Operation Manual SR860 500 kHz DSP Lock in Amplifier ...
Page 5: ...Safety and Preparation For Use iii SR860 DSP Lock in Amplifier ...
Page 6: ...iv Safety and Preparation For Use SR860 DSP Lock in Amplifier ...
Page 54: ...36 Getting Started Chapter 1 SR860 DSP Lock in Amplifier ...
Page 118: ......
Page 172: ......
Page 186: ...168 The FFT Display Appendix B SR860 DSP Lock in Amplifier ...
Page 192: ......