Gazeeka Calibration
1
3
5
2
4
6
With the tractor running:
Make sure the instrument has been on for at least 2
minutes.
Make sure the air path between the antennae is clear
(i.e., no bale or anything else). The end of the bale
should be inside the rear end of the baler doors by at
least a foot.
Press (Item) to select the “ENT for Air Cal” item.
Press (ENT) to select an air path reading to be carried
out.
If no error message displayed, the air path
standardization has been completed OK.
Press (ENT) to go back to the Setup Menu.
Select the Setup Mode – If in Analyze Mode, press
(F1) to select the Setup mode.
Clear the air path and press (F1). Press (ENT) to
calibrate – it will take a few seconds.
Power off then on again.
The instrument is now ready to start measuring bales.