SSI IMMUVIEW READER Quick Manual Download Page 1

ENGLISH (UK)

 QUICK  GUIDE

IMMUVIEW

® 

READER

Instrument for automated 

interpretation of results for 

ImmuView

®

 lateral flow test

Summary of Contents for IMMUVIEW READER

Page 1: ...ENGLISH UK QUICK GUIDE IMMUVIEW READER Instrument for automated interpretation of results for ImmuView lateral flow test ...

Page 2: ...fferent functions of the reader Intended use The ImmuView Reader is an adjunctive instrument for automated reading and interpretation of results for ImmuView lateral flow immunochromatographic in vitro diagnostics test strips Materials provided The following content are supplied with the ImmuView Reader ImmuView Reader 1 x Instrument check cartridge 2 x Strip Carriers Power supply With adaptors fo...

Page 3: ...eader Rear View TFT LCD COLOUR DISPLAY TOUCH SCREEN LED STATUS INDICATOR USB CONNECTION USB CONNECTION ETHERNET CONNECTION CARTRIDGE STRIP CARRIER SLOT STRIP CARRIER INSTRUMENT CHECK CARTRIDGE COMPANY LOGO RATING LABEL AUDIO SPEAKER SERIAL NUMBER UDI LABEL SECURITY LOCK SLOT POWER CONNECTION POWER BUTTON ...

Page 4: ...s is how the startup sequence appears on the ImmuView Reader Hourglass icon appears Loading bar appears and begins to load till full ImmuView logo is displayed with indication of start up progress Main menu Once the start up sequence is completed the first screen you will see is The Home menu screen You are per default logged in as Default User To log in push the key button lower left corner and t...

Page 5: ...est settings please refer to section 8 7 in the user manual Step 2 Press Perform Test on the screen Step 3 Enter Patient ID Enter a unique patient ID using the onscreen keyboard or a scanner Press Accept to save the patient ID NB If you Press X Cancel it cancels the test and test data is not saved A handheld scanner can be attached to the reader We recommend Datalogic QuickScan QD2430 2D ...

Page 6: ...rt to analyse the strip Once you have inserted the Strip Carrier no further actions are required The reading takes around 40 seconds When the reading is done the test result will automatically be shown on the screen The Strip Carrier must be left inserted in the ImmuView Reader during this phase of analysis Result After the analysis the ImmuView Reader makes a sound to indicate that the analysis i...

Page 7: ...in the Strip Carrier section of this Quick Guide and or the ImmuView IFU interpreta tion of result section The ImmuView Reader will automatically save the result which can be accessed in the Results please refer to section 8 12 in the ImmuView Reader user manual for more information To perform another test press this icon To export the test result press this icon Only possible if the ImmuView Read...

Page 8: ...he ImmuView Reader is within the specifications A fail describes that the ImmuView Reader is outside the specifications and the reader is to be put in quarantine and sent to the supplier please refer to the manual Note The ImmuView Reader is by default pre set to prompt the user every 7 days for an Instrument Check LIS integration and other tasks The ImmuView Reader can be connected to the LIS sys...

Page 9: ...ll to ease the removal of the strip Insert strip here THE STRIP CARRIER Add test and wait 15 minutes Add running buffer and whirl gently Sample addition 3 drops 120 µL 2 drops 90 µL Legionella and S pneumoniae positive A B C Legionella and S pneumoniae positive A B C Legionella positive A B C S pneumoniae positive A B C Negative A B C Result interpretation A Control B Legionella C S pneumoniae Loo...

Page 10: ...ont of the Strip Carrier the ImmuView strip is in place and you can now insert the Strip Carrier in the ImmuView Reader Step 4 Insert the Strip Carrier in the ImmuView Reader please see section 5 in the Quick Guide for correct sequence On the reader it will say insert test when it is time to insert the Strip Carrier Back Front Strip placement Make sure P L is visible placed as in the picture NOTIC...

Page 11: ...Strip Carrier Cleaning with 70 Isopropanol and 10 Bleach See manual for further details Quality certificate SSI Diagnostica s development production and sales of in vitro diagnostics are quality assured and certified in accordance with ISO 13485 Information and ordering SSI Diagnostica A S Herredsvejen 2 3400 Hillerød Denmark T 45 4829 9100 support immuview com immuview com 2016 2020 SSI Diagnosti...

Page 12: ...IMMUVIEW COM ...