EVX

®

 

Model 2309

 

Clearing The Airway Is Our #1 Priority 

Operating Instructions 

& Maintenance Manual 

Tel 1+818-504-4054 

Tel 1+800-434-5211 

               , 

INC.

 

11064 Randall Street 
Sun Valley, CA 91352  USA 
 
www.sscor.com 
Email: [email protected] 
[email protected] 

Summary of Contents for EVX 2309

Page 1: ...ng The Airway Is Our 1 Priority Operating Instructions Maintenance Manual Tel 1 818 504 4054 Tel 1 800 434 5211 INC 11064 Randall Street Sun Valley CA 91352 USA www sscor com Email marketing sscor com techsupport sscor com ...

Page 2: ...ister lid will close down when the canister is filled with fluids 8 Install a new canister before testing for vacuum over 300mmHg to minimize the possibility of implosion which can occur when a canister is aged or damaged 9 SSCOR s battery charger charges batteries only The battery charger will not run the pump 10 External equipment intended for connection to signal input signal output or other co...

Page 3: ...of power and can be pre set to be activated immediately upon reaching the distressed patient All controls are clearly labeled and easily accessible The EVX is a portable 12V DC battery operated suction device for resuscitation emergencies in prehospital care The battery is recharged by direct connection to the vehicle or an optional SSCOR fixed voltage charger Model 2309 only A fully charged batte...

Page 4: ...ized Label Attention Consult Accompanying Documents Shipping Carton Manufacturer Shipping Carton Date of Manufacture Shipping Carton Authorized Representative in the European Community Control Panel On Turn Pump On Off Turn Pump Off SSCOR EVX Model 2309 4 Clearing The Airway Is Our 1 Priority Description of Symbols Rev R 12 13 ...

Page 5: ...1 1A into the receptacle 1A This charging method is designed to keep the battery charging at all times If the suction unit is operated while it is hooked up to the vehicle it will utilize the vehicle power and save its own battery for emergency use All SSCOR suction units have a diode to prevent drawdown from the pump to the vehicle electrical system and a fuse to protect the pump from vehicle ele...

Page 6: ...attery quality is suspected to ensure proper performance of the device 1 Check the power source connected LED on the control panel 4 is lit and charge the device overnight to ensure the battery is fully charged 2 Remove the power cord and run the unit from its internal DC battery 3 Check for vacuum by occluding the patient tube and set the vacuum regulator 7 to the maximum vacuum setting 4 Observe...

Page 7: ... REMOVE MAIN CHASSIS SCREWS A 1 screw under the canister holder B 2 screws in the back of the handle B 3 screws in the back of the main chassis C It should not be necessary to open the exhaust door unless fluids have entered the system H DC receptacle Wiring Harness I Battery Bracket J Battery 12V Sealed Lead Acid K N A L Vacuum pump Return to SSCOR or authorized service center for service Do not ...

Page 8: ...acuum pump 6 For technical assistance call 800 434 5211 or international 1 818 504 4054 NOTE No part requires lubrication and lubricants should not be used CHARACTERISTICS SPECIFICATIONS Size 17 L x 9 H x 6 W 43 2cm L x 22 9cm H x 15 2cm W Weight 8 6 pounds 3 9 Kgs Vacuum Pump 12V DC 3 0 A Clinical Airflow 30LPM Exceeds 525mmHg Variable Regulator 7 Controls negative pressure from 50 to 525mmHg Gau...

Page 9: ...istance call 800 434 5211 or 1 818 504 4054 SSCOR EVX Model 2309 9 Clearing The Airway Is Our 1 Priority Maintenance Disinfection As soon as possible after use the single use disposable canister patient tubing and catheter should be discarded according to local regional national requirements for the disposal of hazardous waste materials Clean the exterior of the SSCOR suction unit using a mild det...

Page 10: ...rging Retention Bracket Put the four 1 1 2 screws through the holes in the charging bracket Model 8310 Reattach the washers and nuts to the screws to secure the Charging Retention bracket in place Base Mount part 8314 7 Mark the hole placement for a base mount and use a 5 16 drill to drill the holes Remove the hardware from the hardware packet Coat the screws with Loctite not supplied and place th...