Model 2109

 

Clearing The Airway Is Our #1 Priority 

Operating Instructions 

& Maintenance Manual 

Tel 1+818-504-4054 

Tel 1+800-434-5211 

               , 

INC.

 

11064 Randall Street 
Sun Valley, CA 91352  USA 
 
www.sscor.com 
Email: [email protected] 
[email protected] 

Summary of Contents for 2109

Page 1: ...aring The Airway Is Our 1 Priority Operating Instructions Maintenance Manual Tel 1 818 504 4054 Tel 1 800 434 5211 INC 11064 Randall Street Sun Valley CA 91352 USA www sscor com Email marketing sscor com techsupport sscor com ...

Page 2: ...efore this device is used 5 Do not use in the presence of flammable agents or anesthetics 6 The S SCORT 9 produces a powerful vacuum Do not use the S SCORT 9 to suction neonates 7 Before testing for vacuum over 300mmHg look for an expiration date on the canister where applicable and change the canister if the canister has passed the expiration date to minimize the possibility of implosion which ca...

Page 3: ...quipment To protect the battery after each procedure turn the unit off put the unit on charge and always store the pump at room temperature The only way to assure the battery has functional capacity even if it is indicated the battery is fully charged is to perform the battery check suggested on page 6 of this operations manual The S SCORT 9 is designed to provide instant effective suctioning inde...

Page 4: ...4 Clearing The Airway Is Our 1 Priority Important Features S SCORT 9 Model 2109 Rev L 10 15 ...

Page 5: ...1 Roto molded Chassis 12 Deck Straps Holds Shoulder Strap 13 Tubing Strap Holds Patient Tubing in place 14 Patient Connecting Tubing 9 32 ID SSCOR part 43200 15 Vented Suction Tip HI D Big Stick Large Bore Suction Tip SSCOR part 44241 16 Battery Connection Quick Disconnect 17 Pump Provides Vacuum Source 18 Battery Supplies Electrical Power to Pump 19 Battery Holder Holds Battery in Place 20 Battte...

Page 6: ...C fixed voltage battery charger 21 or a DC power cord for direct connection to the vehicle 23 An active vehicle running calls around the clock will do well to charge the S SCORT 9 from the vehicle s DC charging system The preferred method of charging the S SCORT 9 is to wire the DC power cord to the vehicle DC power system on a properly fused line in front of the master switch This charging method...

Page 7: ...n the Model 2109 is controlled by a two position regulator 10 When fully depressed pushed toward the pump the negative pressure exceeds 525mmHg To reduce negative pressure pull the regulator straight out to the stop In this position the negative pressure will be 120mmHg 15 Attach the patient connecting tube 14 to the patient port 2 on the canister 3 Use 9 32 I D tubing Turn the On Off switch 8 to ...

Page 8: ...er Re secure canister lid Occlude thumb vent with thumb Pump is sluggish Residual materials have collected in the pump head Battery is unable to retain a charge Loose connections Replace pump Replace battery Check connections System shuts down while suctioning heavy particulate matter Vacuum line clogged at canister lid Float valve has closed Remove connector or canister lid and loosen obstruction...

Page 9: ...bilities on SSCOR s part and SSCOR neither assumes nor authorizes any representative or other persons to assume for it any other liability in connection with the sale of SSCOR products This warranty gives you specific legal rights and you may also have other rights that vary from jurisdiction to jurisdiction For countries where minimum warranty terms are determined by statute the warranty term is ...

Page 10: ...istance call 800 434 5211 or 1 818 504 4054 10 Clearing The Airway Is Our 1 Priority Maintenance Disinfection As soon as possible after use the single use disposable canister patient tubing and catheter should be discarded according to local regional national requirements for the disposal of hazardous waste materials Clean the exterior of the SSCOR suction unit using a mild detergent and clear wat...

Page 11: ... will power the suction device for 30 45 minutes Vehicle electrical system powers the pump and charges the battery until disconnected and then the unit switches automatically to its self contained battery AC Charger U L Listed AC DC Dual Mode Battery Charger Switch On Off Rocker Collection Canister 1200cc ml SSCOR part 48041 Patient tubing Vinyl tubing 9 32 ID x 72 L SSCOR part 43200 Suction Tip H...

Page 12: ... the new battery in the battery holder reconnect the Molex connection and secure the battery using the Velcro strap Turn the device on and verify negative pressure from the vacuum tubing Re attach the battery plate to the base plate Dispose of the old battery according to local regional national requirements for the disposal of sealed lead acid batteries 12 Clearing The Airway Is Our 1 Priority Re...