Model SR735
i
-KG Wheelchair Scale
Operating and Service Manual
Part No. MAN735
i
-KG_MICROBAN_201208
Page 9 of 20
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
Copyright 2020
S
Instruments, Inc.
BUTTON FUNCTIONS
ZERO / WEIGH
Press and hold to zero. The display will read “
WT = 0.0 Kg
”. This is used to zero the
system before placing the patient on the scale system. This action also resets previously
stored weight, height and BMI values to zero. Ensure that nothing is in contact with the
weighing surface during this procedure.
Press to weigh. Weight stable indicator “
” flashes then remains solid when stable. Auto
stores stable weight in memory.
SEND (PRINTER / EHR)
Press to send stored values to EHR or to printer. Output values include time, date, weight,
height, and BMI. Also, indicated on display when paper is out, “
PAPER OUT
” and when
door is open, “
PAPER DOOR OPEN
”.
RECALL
Press to recall the last stable weight. Display will scroll thru “
WT
”, “
HT
”, and “
BMI
”
stored values in approximately two second intervals.
ENTER
Press to save height data and display BMI calculation. Stable weight, height and BMI
values are stored in memory until next stable weight is stored or until system is zeroed.
BMI
Press to calculate the BMI. When the “
BMI
” is pressed, the default starting value “
HT =
165 cm
” is displayed. If there is no stored stable weight, the display will indicate “
NO
WEIGHT DATA
” and then go back to the weigh screen “
WT = 0.0 Kg
”.
Figure 6: Button Display
Continued next page