BUTTON FUNCTIONS con’t
ENTER
Press to save change in digits for calibration, for unit’s set-up, for time and date set-up
and saving completed calibration data.
UP / SELECT
Press
UP
to adjust height up from the default for BMI calculation, to increase the scale’s
“on time”, or to select a digit when setting time and date.
DOWN / CHANGE
Press
DOWN
to adjust the height down from the default for BMI calculation, to
decrease the scale’s “on time”, or to change the value of a selected digit when setting
time and date.
BASIC SYSTEM OPERATION
SETTING SYSTEM ZERO
Make sure the scale is free and clear of any obstructions. Press and hold the
ZERO /
WEIGH
button. The displayed message will indicate
“
HOLD TO ZERO
” and count
down to zero. Release the button when display message indicates “
HANDS OFF
”.
Make
sure that nothing is in contact with the scale while zeroing the system. In a few seconds, the
display will read “
WT = 0.0 Lb
”
(or
Kg
)
.
WEIGHING
Position the patient on the scale. The weight stable indicator
“
” flashes on the display.
When the weight is stable, the weight stable indicator remains solid. The display will
indicate the patient’s weight in either pounds or kilograms; example: “
WT = 123.5 Lb
”.
The stable weight is auto stored in memory.
RECALLING LAST STORED WEIGHT
Press to recall last stored weight. The stored weight will be erased when the scale is zero
or another stable weight is stored.