Model SR416i-KG - Patient Floor Scale 
Operating and Service Manual 
Part No. MAN416I-KG_200611 

 

Page 3 of 16 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email: 

[email protected]

     

  

Copyright 2020

 

S

Instruments, Inc. 

PACKING CHECKLIST – Model SR416

i

-KG 

Patient Floor Scale 

 

ASSEMBLY 

 

STEP 1

: Unpack the scale system and check parts against the 

PACKING CHECKLIST

. If there 

are any missing or damaged parts, call the Service Hotline at:  1-800-654-6360. 

STEP 2

: Place the SR416

i

-KG

 

scale system on a flat, hard floor surface. Ensure all four corners of 

scale are in contact with the floor and scale does not rock

NOTE:

 Scale should not be placed on carpet or uneven floor surfaces as this may interfere with 

scale accuracy. 

STEP 3

: Connect the power cable to the scale and plug the power adapter into a wall outlet. 

 

SYSTEM DESCRIPTION and INTENDED USE 

 

SYSTEM DESCRIPTION 

The SR416

i

-KG Patient Floor Scale employs the latest in microprocessor and load cell technology 

to  provide  accurate  and  repeatable  weight  data.  Four  (4)  identically  matched  transducers  are 
strategically placed to ensure an accurate representation of the patient’s weight. 

The patient’s weight is displayed on a LCD screen. With a push of a button, weight data may be 
viewed in kilograms. 

INTENDED USE 

The SR416

i

-KG Patient Floor Scale is a unique lightweight, low profile stand-on scale system for 

weighing  ambulatory  patients.  The  wide  platform  is  designed  to  easily  accommodate  larger 
patients. 

The  SR416

i

-KG  Patient  Floor Scale  is  a preferred means of  gathering patient weight data of patients 

weighing up to 400 kilograms. 

 

 

DESCRIPTION 

QUANTITY 

 

SCALE BASE w/BUILT IN DISPLAY 

1 ea. 

 

AC WALL POWER ADAPTER (MEDICAL GRADE) 

1 ea. 

 

CERTIFICATE OF CALIBRATION 

1 ea. 

 

WARRANTY CARD 

1 ea. 

 

MANUAL 

1 ea. 

WARNING 

DO NOT EXCEED  

MAXIMUM WEIGHT LIMIT OF 

400 KG 

Summary of Contents for SR416i-KG

Page 1: ...KG_200611 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S I...

Page 2: ...struments Inc TABLE OF CONTENTS PACKING CHECKLIST MODEL SR416I KG 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 3 STORAGE AND TRANSPORTATION 4 MICROBAN ANTIMICROBIAL PROTECTION 4 CLEANING AND DISIN...

Page 3: ...power adapter into a wall outlet SYSTEM DESCRIPTION and INTENDED USE SYSTEM DESCRIPTION The SR416i KG Patient Floor Scale employs the latest in microprocessor and load cell technology to provide accu...

Page 4: ...TATION To maintain proper operation of this instrument storage and transport conditions should not vary outside the following conditions Relative Humidity Range of 0 to 85 Ambient Temperature Range of...

Page 5: ...eas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disin...

Page 6: ...ibrated range AUTO ZERO One button operation AUTO WEIGH Scale will automatically turn on and lock in on a stable weight AUTO POWER DOWN After 1 minute AVERAGING Automatic digital filter POWER SUPPLY M...

Page 7: ...play reads ZERO then release and wait until the display reads 0 0 When the patient is in position on the scale the weight will be displayed NOTE When the system is zeroed the Hold Recall memory is era...

Page 8: ...rnatively press the Zero Weigh button and wait for the display to indicate ZERO Release the Zero Weigh button and the display will read 0 0 Position the patient on the platform Weight will be displaye...

Page 9: ...by the operational amplifier The output of the operational amplifier is digitized by the analog to digital converter The converter integrates the analog signal onto the integrating capacitor over a sh...

Page 10: ...et in calibration mode Both procedures are found below To access all settings below remove the feet spacers and seven 7 screws holding the bottom cover on Press the calibration button Figure 2 and CAL...

Page 11: ...n of lb or kg with mode button sec Set Auto Power Down Timer between 15 and 900 seconds grav Set Gravitational Acceleration in m s2 Full Applied calibrated mass at full scale Half Used with three poin...

Page 12: ...cceleration from the CAL menu by pressing the Left Button when the display shows grav b The display will show only the four digits to the right of the decimal point 9 XXXX c Set the local gravitationa...

Page 13: ...t by pressing the Left Button to quit without saving changes e Press the Left Button again to save and the display will show SAV d Calibrate using the Two Point Calibration Curve zero and full scale N...

Page 14: ...ECTIVE ACTION Weight reading is much lower than expected Check that the platform is clear underneath stand on each corner to see if one corner is not weighing correctly System fails to perform correct...

Page 15: ...liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In n...

Page 16: ...rt No MAN416I KG_200611 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstrument...

Reviews: