SR Scales SR416i-KG Operating And Service Manual Download Page 10

Model SR416i-KG - Patient Floor Scale 
Operating and Service Manual 
Part No. MAN416I-KG_200611 

 

Page 10 of 16 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email: 

[email protected]

     

  

Copyright 2020

 

S

Instruments, Inc. 

CALIBRATION 

 
 
 
 
 
 
 
 
 
 
 

INITIAL SYSTEM SETUP 

When  initially  set  up,  calibration  is  factory  set  and  re-calibration  is  not  needed.  The  Local 
Gravitational  Acceleration  may  have  to  be  re-set  for  the  current  geographical  location  for 
improved accuracy. Automatic Shut Off is shipped with a factory default of 60 seconds. If a longer 
period  of  time  (up  to  900  seconds)  is  desired,  then  it  will  need  to  be  re-set  in  calibration  mode. 
Both procedures are found below. 

To access all settings below, remove the feet, spacers and seven (7) screws holding the bottom cover on. 
Press the 

calibration button 

(Figure 2) and 

CAL mode will now be active. The LCD will show “CAL”.

 

 
 
 
 
 
 
 
 
 
 
 
 
 

Figure 2: Calibration Button

 

 

CAUTION 

The  integrated  circuits  and  semiconductors 
on  the  printed  circuit  boards  may  be 
damaged by electrostatic discharge (ESD). Be 
sure to use proper handling precautions at all 
times.  

IMPORTANT 

CALIBRATION  CHECK:  Qualified  service  personnel  only  should  perform  this 
procedure. Load cells have no user serviceable components and should not be tampered 
with  for  any  reason.  Re-calibration  is  generally  not  required,  but  should  be  verified 
periodically  to  ensure  accuracy.  The  recommendation  for  calibration  check  is  at  least 
once every 12 months, or as individual maintenance policy requires. 

 

Summary of Contents for SR416i-KG

Page 1: ...KG_200611 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S I...

Page 2: ...struments Inc TABLE OF CONTENTS PACKING CHECKLIST MODEL SR416I KG 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 3 STORAGE AND TRANSPORTATION 4 MICROBAN ANTIMICROBIAL PROTECTION 4 CLEANING AND DISIN...

Page 3: ...power adapter into a wall outlet SYSTEM DESCRIPTION and INTENDED USE SYSTEM DESCRIPTION The SR416i KG Patient Floor Scale employs the latest in microprocessor and load cell technology to provide accu...

Page 4: ...TATION To maintain proper operation of this instrument storage and transport conditions should not vary outside the following conditions Relative Humidity Range of 0 to 85 Ambient Temperature Range of...

Page 5: ...eas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disin...

Page 6: ...ibrated range AUTO ZERO One button operation AUTO WEIGH Scale will automatically turn on and lock in on a stable weight AUTO POWER DOWN After 1 minute AVERAGING Automatic digital filter POWER SUPPLY M...

Page 7: ...play reads ZERO then release and wait until the display reads 0 0 When the patient is in position on the scale the weight will be displayed NOTE When the system is zeroed the Hold Recall memory is era...

Page 8: ...rnatively press the Zero Weigh button and wait for the display to indicate ZERO Release the Zero Weigh button and the display will read 0 0 Position the patient on the platform Weight will be displaye...

Page 9: ...by the operational amplifier The output of the operational amplifier is digitized by the analog to digital converter The converter integrates the analog signal onto the integrating capacitor over a sh...

Page 10: ...et in calibration mode Both procedures are found below To access all settings below remove the feet spacers and seven 7 screws holding the bottom cover on Press the calibration button Figure 2 and CAL...

Page 11: ...n of lb or kg with mode button sec Set Auto Power Down Timer between 15 and 900 seconds grav Set Gravitational Acceleration in m s2 Full Applied calibrated mass at full scale Half Used with three poin...

Page 12: ...cceleration from the CAL menu by pressing the Left Button when the display shows grav b The display will show only the four digits to the right of the decimal point 9 XXXX c Set the local gravitationa...

Page 13: ...t by pressing the Left Button to quit without saving changes e Press the Left Button again to save and the display will show SAV d Calibrate using the Two Point Calibration Curve zero and full scale N...

Page 14: ...ECTIVE ACTION Weight reading is much lower than expected Check that the platform is clear underneath stand on each corner to see if one corner is not weighing correctly System fails to perform correct...

Page 15: ...liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In n...

Page 16: ...rt No MAN416I KG_200611 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstrument...

Reviews: