background image

Model SR416i - Patient Floor Scale 
Operating and Service Manual 
Part No. MAN416I_200611 

 

Page 1 of 16 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email: 

[email protected]

     

  

Copyright 2020

 

S

Instruments, Inc. 

 
 

S

 

by 

Instruments, Inc.

 

 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Operating and Service 

Manual 

Model 

SR416

 

Patient Floor Scale 

Summary of Contents for Model SR416i

Page 1: ...6I_200611 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S In...

Page 2: ...struments Inc TABLE OF CONTENTS PACKING CHECKLIST MODEL SR416I 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 3 STORAGE AND TRANSPORTATION 4 MICROBAN ANTIMICROBIAL PROTECTION 4 CLEANING AND DISINFEC...

Page 3: ...YSTEM DESCRIPTION and INTENDED USE SYSTEM DESCRIPTION The SR416i Patient Floor Scale employs the latest in microprocessor and load cell technology to provide accurate and repeatable weight data Four 4...

Page 4: ...ION To maintain proper operation of this instrument storage and transport conditions should not vary outside the following conditions Relative Humidity Range of 0 to 85 Ambient Temperature Range of 14...

Page 5: ...Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disinfec...

Page 6: ...calibrated range AUTO ZERO One button operation AUTO WEIGH Scale will automatically turn on and lock in on a stable weight AUTO POWER DOWN After 1 minute AVERAGING Automatic digital filter POWER SUPP...

Page 7: ...is in position on the scale the weight will be displayed NOTE When the system is zeroed the Hold Recall memory is erased LB KG BUTTON Patient weight data may be viewed in either pounds or kilograms P...

Page 8: ...ively press the Zero Weigh button and wait for the display to indicate ZERO Release the Zero Weigh button and the display will read 0 0 Position the patient on the platform Weight will be displayed in...

Page 9: ...the operational amplifier The output of the operational amplifier is digitized by the analog to digital converter The converter integrates the analog signal onto the integrating capacitor over a short...

Page 10: ...in calibration mode Both procedures are found below To access all settings below remove the feet spacers and seven 7 screws holding the bottom cover on Press the calibration button Figure 2 and CAL mo...

Page 11: ...lb or kg with mode button sec Set Auto Power Down Timer between 15 and 900 seconds grav Set Gravitational Acceleration in m s2 Full Applied calibrated mass at full scale Half Used with three point ca...

Page 12: ...on from the CAL menu by pressing the Left Button when the display shows grav b The display will show only the four digits to the right of the decimal point 9 XXXX c Set the local gravitational acceler...

Page 13: ...y pressing the Left Button to quit without saving changes e Press the Left Button again to save and the display will show SAV d Calibrate using the Two Point Calibration Curve zero and full scale NOTE...

Page 14: ...IVE ACTION Weight reading is much lower than expected Check that the platform is clear underneath stand on each corner to see if one corner is not weighing correctly System fails to perform correctly...

Page 15: ...bilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no e...

Page 16: ...rt No MAN416I_200611 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments I...