CALIBRATION (CONT’D) 

 

STEP 5:

   Press  button  “

S4

”  to  scroll  through  menu  options  until  “

FULL

”  is  displayed.  Press  the 

ZERO/WEIGH

 button. 

STEP 6: 

Set the “

FULL

” value to the set of selected weights from Step 1. Use the 

ZERO/WEIGH

 

button to select digit positions and use the 

S4

 button to change the value (Figure 5). 

STEP 7:

  When finished, the display will read “

SAVE

”. Press the 

ZERO/WEIGH

 button to save, or 

press the 

S4 

button then 

ZERO/WEIGH

 to “

QUIT

”. 

STEP 8:

  Press  the 

S4

  button  to  scroll  to  the  menu  option  “

2  PT

”  and  press  the 

ZERO/WEIGH

 

button. 

STEP 9:

  Ensure nothing is in contact with the scale, then press the 

ZERO/WEIGH

 button when the 

display reads “

ZERO

”. 

STEP 10:

  When the display reads “

FULL

”, place the weight from Step 1 on the platform and press the 

ZERO/WEIGH

 button. 

STEP 11:

  Press the 

ZERO/WEIGH

 button once more to “

SAVE”

. To exit without saving, press the 

S4 

button

 

to select “

QUIT

”, then press the 

ZERO/WEIGH

 button.

 

STEP 12:

  When finished making adjustments, remove all weights and top platform from the scale. 

STEP 13: 

Replace  the  display  cover  and  securely  tighten  screws.  Re-attach  display  to  the  frame  and 

securely tighten screws. Place the top platform back on frame. 

NOTE: TIGHTEN SCREWS ONLY UNTIL RUBBER SEAL STARTS TO COMPRESS

 

OVERTIGHENING MAY CAUSE WARPAGE AND GAPS 

STEP 14:

  Press the 

ZERO/WEIGH

 button to zero the scale. When the display indicates “

0.0

”, place a 

calibrated weight on the scale again and check against the weight displayed. If the weight is not correct, 
recalibrate. (If calibration cannot be accomplished, call the Service Department.) 

 
 
 
 
 
 
 

TROUBLESHOOTING 

 

SYMPTOM 

REASON/CORRECTIVE ACTION 

Weight reading is much lower than 
expected. 

Check  that  the  scale  is  clean  underneath,  stand  on 
each  of  the  four  corners  to  see  if  one  corner  is  not 
weighing correctly. 

For additional information or assistance, telephone our Service Hotline: 1-800-654-6360 

or e-mail: 

[email protected]

 

CAUTION 

The integrated circuits and semiconductors on 
the printed circuit boards may be damaged by 
electrostatic  discharge  (ESD).  Be  sure  to  use 
proper handling precautions at all times.  

Summary of Contents for SR Scales SRV715

Page 1: ...SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2019 SInstruments Inc S by SR Instruments Inc Operatin...

Page 2: ...CHECKLIST MODEL SRV715 3 REPLACEMENT PARTS AND ACCESSORIES 5 SYSTEM DESCRIPTION AND INTENDED USE 5 MAINTENANCE AND CLEANING 6 STORAGE AND TRANSPORTATION 6 SPECIFICATIONS 7 BUTTON FUNCTIONS 8 BASIC SY...

Page 3: ...cales com email sri srinstruments com Copyright 2019 SInstruments Inc PACKING CHECKLIST Model SRV715 Platform Scale DESCRIPTION QUANTITY TOP PANEL 44 in 112 cm x 72 in 183 cm 1 ea BOTTOM FRAME 48 in 1...

Page 4: ...x 6 AA cell batteries according to BATTERY REPLACEMENT instructions Remove the four 4 screws from the metal cover on the display located under the bottom frame Remove cover and insert batteries Attach...

Page 5: ...form Scale derives its power from six 6 AA cell batteries Weight is displayed with a display resolution of 1 pound or 0 5 kilograms The SRV715 Medium Animal Platform Scale is equipped with a Bluetooth...

Page 6: ...cratched by abrasive cleaners Waterproof seals can also be damaged We recommend mild soap and water for general cleaning and disinfecting STORAGE and TRANSPORTATION STORAGE If storing this equipment f...

Page 7: ...BLUETOOTH ON TIME 10 minutes from last button press AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA cell batteries CALIBRATION Calibration is traceable to NIST standards OPERATING CONDITIONS...

Page 8: ...e the free SR Scales app Google Play store to connect with the scale from any Android device to allow for remote wireless operation Use your Android device to display weight and control the SRV715 sca...

Page 9: ...correct interface connection Ensure that CONNECTED is displayed on the remote wireless display device Press the ZERO button on the display The display will read 0 0 and be ready for weighing Press PO...

Page 10: ...l turn over the Base Assembly remove the four 4 screws from the metal cover on the display and remove the battery cover and gasket STEP 2 Remove and replace the six 6 AA cell batteries STEP 3 Replace...

Page 11: ...our strain gauges mounted in a full Wheatstone bridge configuration These bridges convert the physical movement of the force cell due to the applied mass on the system into minute changes in electrica...

Page 12: ...te scale reading Remove weights STEP 4 The scale reading should be within the Calibration Tolerance Table Figure 4 SETTING CALIBRATION STEP 1 Select known calibrated weights traceable to NIST NOTE Cal...

Page 13: ...to SAVE To exit without saving press the S4 button to select QUIT then press the ZERO WEIGH button STEP 12 When finished making adjustments remove all weights and top platform from the scale STEP 13...

Page 14: ...and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be liab...

Page 15: ...and Service Manual Part No MAN715_190719 Page 15 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyrigh...

Page 16: ...AN715_190719 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2019 SInstruments Inc S By...

Reviews: