BASIC SYSTEM OPERATION 

 

Ensure that the scale is free and clear of any obstructions before operating. 

Press  the  “

DISPLAY

”  button  and  hold  it  for  2-3  seconds.    The  display  will  read  “0.0”  and  be 

ready for weighing. 

When a mat or pad is to be used on the scale platform, the weight of the pad must be tared out.  
Place the mat on the platform, then press and hold the “

DISPLAY

” button for 2-3 seconds.  The 

display will read “0.0”. 
 

Note

The display will automatically shut down after five minutes.  The zero (TARE) 

 

 

will be stored in memory. 

 

Assist the animal on the scale platform; its weight will appear on the display. 
 

Note

The display will automatically shut down after five minutes.  Press the 

 

 

DISPLAY

” button to restart. 

 

It is recommended that the system be zeroed prior to each subject being weighed. 

 

BATTERY REPLACEMENT 

 

 

Figure 4: Battery Replacement 

 

REPLACING BATTERIES 

 

When battery replacement is needed, an indicator will appear on the display.   

STEP 1: 

Turn the platform over and remove the four screws on the Battery Cover.   

STEP 2: 

Remove and replace the six (6) “AA” cell batteries.  Replace the screws in the Battery 
Cover. 

NOTE: TIGHTEN SCREWS ONLY UNTIL RUBBER SEAL STARTS TO COMPRESS 

STEP 3:

  Press the “

DISPLAY

” button to confirm display is working. 

STEP 4: 

Position the scale back into weighing position.  

STEP 5:

  Zero the system.  Scale is ready for use. 

Battery Cover 

Summary of Contents for SR Scales SRV712W

Page 1: ...f 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2017 SInstruments Inc S by SR Instruments Inc Ope...

Page 2: ...CHECKLIST MODEL SRV712W 3 ASSEMBLY 4 REPLACEMENT PARTS AND ACCESSORIES 5 SYSTEM DESCRIPTION AND INTENDED USE 5 MAINTENANCE AND CLEANING 5 STORAGE AND TRANSPORTATION 6 SPECIFICATIONS 7 BUTTON FUNCTIONS...

Page 3: ...Y 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2017 SInstruments Inc PACKING CHECKLIST Model SRV712W Marine Platform Scale DESCRIPTION QUANTITY PL...

Page 4: ...rine Scale Platform 1 and check parts against the PACKING CHECKLIST If there are any missing or damaged parts call the Service Hotline at 1 800 654 6360 STEP 2 Place the Platform upside down on the fl...

Page 5: ...accurate representation of the subject s weight The SRV712W Marine Platform Scale derives its power from six 6 each AA batteries With the push of a button weight is displayed with a displayed resolut...

Page 6: ...asive cleaners The waterproof seal can also be damaged We recommend mild soap and water for general cleaning and disinfecting STORAGE and TRANSPORTATION STORAGE If storing this equipment for periods l...

Page 7: ...dle BLUETOOTH ON TIME 5 minutes from last button press AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA batteries CALIBRATION Calibration is traceable to NIST standards OPERATING CONDITIONS No...

Page 8: ...lay box on the underside of the scale Select the mode for weighing with this Toggle Button SR SCALES APP Use the free SR Scales app Google Play store to connect with the scale from any Android device...

Page 9: ...atically shut down after five minutes The zero TARE will be stored in memory Assist the animal on the scale platform its weight will appear on the display Note The display will automatically shut down...

Page 10: ...four strain gauges mounted in a full Wheatstone bridge configuration These bridges convert the physical movement of the force cell due to the applied mass on the system into minute changes in electric...

Page 11: ...selected weight from Step 1 Use the DISPLAY button to select digit positions and use button S4 to change the value STEP 7 When finished the display will read SAVE Press the DISPLAY button to save or p...

Page 12: ...urely tighten screws NOTE TIGHTEN SCREWS ONLY UNTIL RUBBER SEAL STARTS TO COMPRESS STEP 14 Press and hold the DISPLAY button to zero the platform When the display indicates 0 0 place the calibrated we...

Page 13: ...t and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be lia...

Page 14: ...AN712W_170816 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2017 SInstruments Inc S By...

Reviews: