CALIBRATION (Cont’d) 

 

SETTING CALIBRATION: 

 
 
 
 
 
 
 
 
 
 
 
 

NOTE:

 Ensure that nothing is in contact with the scale system during this procedure. Remove hands from 

the system when noting the displayed calibration results. 
 

STEP 1:

  Remove the two (2) screws on the end cap closest to the display. Put the scale system into the 

Calibration Mode by switching the calibration switch on the display board (Figure 5). 

STEP  2:

    Select  known  calibrated  weights, traceable  to  NIST,  up  to the  Full Scale  set  value (maximum 

capacity). 

STEP 3:

  Press the 

MENU

 button until “

FULL 500.0 lb

” is displayed. Set the 

FULL

 value to the actual 

quantity of calibrated weight being used for Full Scale. Use the 

UP

 arrow button to select the digit and the 

DOWN

 arrow button to change digits. Press 

ENTER

 to save changes. Press 

WEIGH

 button to abort any 

changes. 

STEP 4:

  Press the 

MENU

 button until “

HALF 250.0 lb

” is displayed. Set the 

HALF

 value to the actual 

quantity of calibrated weight being used for Half Scale. Use the 

UP

 arrow button to select the digit and the 

DOWN

 arrow button to change digits. Press 

ENTER

 to save changes. Press 

WEIGH

 button to abort any 

changes. 

NOTE

:  The Half-Scale value is a value between zero and the Full Scale values. It is usually close to half 

the Full Scale value. 

STEP 5:

  Press 

MENU

 button until “

3 PT CAL

” is displayed. Press the 

UP

 arrow button to select. 

STEP 6:  

Remove all weight from the platform when the display prompts to “

ZERO THE SCALE

” and 

press the 

UP 

arrow button to save. 

STEP 7:

  Place the 

HALF 

scale calibrated weights on platform when the display prompts to “

ADD HALF 

SCALE

”. Allow weight to stabilize. Press the 

UP

 arrow button to save. 

STEP 8:

  Place the 

FULL 

scale calibrated weights on platform when the display prompts to “

ADD FULL 

SCALE

”. Allow weight to stabilize. Press the 

UP 

arrow button to save. 

STEP 9

:  Press the 

ENTER 

button when the display prompts to “

SAVE CALIBRATION?

” to save the 

calibration or press the 

WEIGH

 button

 

to exit without saving

.

 

STEP 10:

  Switch the scale system out of the Calibration Mode on the display board (Figure 5). 

STEP 11:

  Reattach the end cap with the two (2) screws. 

Figure 5:  Calibration Switch Location 

CALIBRATION 

SWITCH

 

Summary of Contents for SR Scales SR7005i

Page 1: ...00520 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instru...

Page 2: ...F CONTENTS PACKING CHECKLIST 3 INSTALLATION 3 MICROBAN ANTIMICROBIAL PROTECTION 5 SYSTEM DESCRIPTION AND INTENDED USE 6 CLEANING AND DISINFECTING 7 STORAGE 7 SPECIFICATIONS 8 BUTTON FUNCTIONS 9 BASIC...

Page 3: ...tends 23 into the room hallway when folded down Hallway must be wide enough to allow suitable access and egress for personnel using the hallway If using the AC power supply a wall outlet must be locat...

Page 4: ...walls use a stud finder and locate any studs in the mounting area STEP b Measure up from the floor 27 5 8 and make a mark at a stud location This will be the location of bottom edge of the wall mount...

Page 5: ...ons STEP b For stud wall mounting locations without a stud use a screwdriver to drive the provided self drilling wall anchors into the wall It is not recommended to use a power drill to set the wall a...

Page 6: ...t s weight is displayed on a 16 character dot matrix LCD w LED backlight The weight data may be viewed either in pounds or kilograms with a displayed resolution of 0 1 for each The SR7005i Wall Mount...

Page 7: ...sodium hypochlorite or 70 isopropyl alcohol are suitable for display user interface and other scale contact surfaces After disinfecting use a soft cloth dampened with clean water and dry with a soft...

Page 8: ...AUTO POWER DOWN Adjustable between 30 to 300 seconds AVERAGING Automatic digital filter POWER SUPPLY Six 6 C cell batteries Replace battery indicator on Display Medical grade AC power supply CALIBRAT...

Page 9: ...last stored weight The stored weight will be erased when the scale is zeroed or another stable weight is stored BMI Press to calculate BMI When the BMI is pressed the default value HT 65 in or HT 165...

Page 10: ...ng time and date BASIC SYSTEM OPERATION SETTING SYSTEM ZERO Scale will auto zero when the platform is first lowered Make sure the scale is free and clear of any obstructions When platform is left down...

Page 11: ...lied mass on the system into minute changes in electrical resistance These changes in resistance produce a voltage difference across the Wheatstone bridge which is amplified by the operational amplifi...

Page 12: ...NIST NOTE DO NOT USE barbells or uncalibrated weights STEP 2 Zero the scale by pressing and holding the ZERO button STEP 3 Place the calibrated weights on the scale Wait for scale to stabilize note sc...

Page 13: ...button until HALF 250 0 lb is displayed Set the HALF value to the actual quantity of calibrated weight being used for Half Scale Use the UP arrow button to select the digit and the DOWN arrow button t...

Page 14: ...attery Wait five seconds then re install battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure the membrane switch label is c...

Page 15: ...liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In n...

Page 16: ...al Part No MAN7005i_200520 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstrum...

Reviews: