BATTERY REPLACEMENT 

 

STEP 1

:  The display will read “REPLACE BATTERY”. 

STEP 2

:  Unscrew the two (2) battery cover screws and remove the battery compartment cover.

 

STEP 3

:  Remove and replace ALL six (6) “C” cell batteries. Refer to battery holder for placement 

orientation.

 

STEP 4

:  Press the “WEIGH” button to confirm that “REPLACE BATTERY” is not displayed. 

STEP 5

:  Replace cover and tighten the two (2) cover screws.

 

STEP 6

:  Zero the system.

 

 
 
 

THEORY OF OPERATION 

 
SR Instruments patient weighing systems are digital scales. Strain-gauge force cells convert the force of an 
applied weight into an analog signal. This signal is amplified by an operational amplifier and converted to a 
digital  signal  by  an  on-chip  analog  to  digital  converter.  The  digital  signal  is  filtered,  converted  to 
appropriate units, and displayed on the liquid crystal display. 
 
Strain-gauge force cells each contain four strain gauges mounted in a full Wheatstone-bridge configuration. 
These bridges convert the physical movement of the force cell, due to the applied mass on the system, into 
minute changes in electrical resistance. These changes in resistance produce a voltage difference across the 
Wheatstone-bridge, which is amplified by the operational amplifier. The amplifier is configured to current 
sum the output of each cell, with potentiometers serving to normalize the sensitivity (voltage out per unit of 
weight applied) of each bridge. The offset potentiometer produces a small current, which nulls the output of 
the amplifier for an unloaded system. 
 
The  output  of  the  operational  amplifier  is  digitized  by  the  analog  to  digital  converter.  The  sigma-delta 
converter  sums  a  rapid  sequence  of  0's  (0  volts)  and  1's  (reference  voltage)  to  achieve  balance  with  the 
input signal from the amplifier. 
 
The micro-controller filters the digital output of the analog to digital converter, subtracts the value saved 
during  the  system  zero  operation and  scales the  filtered  output, and then displays  the result  on the liquid 
crystal  display.  The  micro-controller  performs  a  moving-median  filter  of  data  for  continuous  weigh  the 
micro-controller performs checks for signal stability before locking in on the reading. 
 
The  micro-controller  can  be  placed  in  a  calibration  mode,  where  the  system  can  be  re-calibrated.  In  the 
calibration  mode,  the  system  slope  is  calculated  from  two  points  (zero  and  full  scale)  in  the  2-point 
calibration mode or the slope and change in slope is calculated from three points (zero, half, and full scale) 
in the 3-point calibration mode. 
 

Summary of Contents for SR Scales SR7005i

Page 1: ...00520 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instru...

Page 2: ...F CONTENTS PACKING CHECKLIST 3 INSTALLATION 3 MICROBAN ANTIMICROBIAL PROTECTION 5 SYSTEM DESCRIPTION AND INTENDED USE 6 CLEANING AND DISINFECTING 7 STORAGE 7 SPECIFICATIONS 8 BUTTON FUNCTIONS 9 BASIC...

Page 3: ...tends 23 into the room hallway when folded down Hallway must be wide enough to allow suitable access and egress for personnel using the hallway If using the AC power supply a wall outlet must be locat...

Page 4: ...walls use a stud finder and locate any studs in the mounting area STEP b Measure up from the floor 27 5 8 and make a mark at a stud location This will be the location of bottom edge of the wall mount...

Page 5: ...ons STEP b For stud wall mounting locations without a stud use a screwdriver to drive the provided self drilling wall anchors into the wall It is not recommended to use a power drill to set the wall a...

Page 6: ...t s weight is displayed on a 16 character dot matrix LCD w LED backlight The weight data may be viewed either in pounds or kilograms with a displayed resolution of 0 1 for each The SR7005i Wall Mount...

Page 7: ...sodium hypochlorite or 70 isopropyl alcohol are suitable for display user interface and other scale contact surfaces After disinfecting use a soft cloth dampened with clean water and dry with a soft...

Page 8: ...AUTO POWER DOWN Adjustable between 30 to 300 seconds AVERAGING Automatic digital filter POWER SUPPLY Six 6 C cell batteries Replace battery indicator on Display Medical grade AC power supply CALIBRAT...

Page 9: ...last stored weight The stored weight will be erased when the scale is zeroed or another stable weight is stored BMI Press to calculate BMI When the BMI is pressed the default value HT 65 in or HT 165...

Page 10: ...ng time and date BASIC SYSTEM OPERATION SETTING SYSTEM ZERO Scale will auto zero when the platform is first lowered Make sure the scale is free and clear of any obstructions When platform is left down...

Page 11: ...lied mass on the system into minute changes in electrical resistance These changes in resistance produce a voltage difference across the Wheatstone bridge which is amplified by the operational amplifi...

Page 12: ...NIST NOTE DO NOT USE barbells or uncalibrated weights STEP 2 Zero the scale by pressing and holding the ZERO button STEP 3 Place the calibrated weights on the scale Wait for scale to stabilize note sc...

Page 13: ...button until HALF 250 0 lb is displayed Set the HALF value to the actual quantity of calibrated weight being used for Half Scale Use the UP arrow button to select the digit and the DOWN arrow button t...

Page 14: ...attery Wait five seconds then re install battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure the membrane switch label is c...

Page 15: ...liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In n...

Page 16: ...al Part No MAN7005i_200520 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstrum...

Reviews: