Model SR7005
i
Wall Mount In-Room Mini Scale
Operating and Service Manual
Part No. MAN7005i_200520
Page 10 of 16
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
Copyright 2020
S
Instruments, Inc.
BUTTON FUNCTIONS (Cont’d)
ENTER
Press to save change in digits for calibration, for unit’s set-up, for time and date set-up and
saving completed calibration data.
UP / SELECT
Press
UP
to adjust height up from the default for BMI calculation, to increase the scale’s “on
time” or to select a digit when setting time and date.
DOWN / CHANGE
Press
DOWN
to adjust the height down from the default for BMI calculation, to decrease the
scale’s “on time”, or to change the value of a selected digit when setting time and date.
BASIC SYSTEM OPERATION
SETTING SYSTEM ZERO
Scale will auto-zero when the platform is first lowered. Make sure the scale is free and clear of any
obstructions. When platform is left down, press and hold the
ZERO / WEIGH
button. The displayed
message will indicate “
HOLD TO ZERO
” and count down to zero. Release the button when
display message indicates “
HANDS OFF
”. Make sure that nothing is in contact with the scale while
zeroing the system. In a few seconds, the display will read “
WT = 0.0 Lb
” (or
0.0 Kg
).
WEIGHING
Position the patient on the scale. The weight stable indicator “
” flashes on the display. When the
weight is stable, the weight stable indicator remains solid. The display will indicate the patient’s
weight in either pounds or kilograms; example: “
WT = 123.5 Lb
”. The stable weight is auto
stored in memory.
RECALLING LAST STORED WEIGHT
Press to recall last stored weight. The stored weight will be erased when the scale is zeroed or
another stable weight is stored.