SR Instruments SR Scales SR630i Operating And Service Manual Download Page 9

Model:  SR630

i

-KG Pediatric Scale 

Operating and Service Manual  
Part No.:  MAN630

i

-KG_200915 

Page 9 of 20 

 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL

www.srscales.com

 

email: 

[email protected]

          Copyright 2020

 

S

Instruments, Inc. 

MAINTENANCE and CLEANING 

 

Exercise caution when cleaning the display window and pediatric cradle as both can be scratched 
by abrasive cleaners.  Mild soap and water is recommended for general cleaning. 

CLEANING 

To clean the pediatric cradle and display / user interface: 

 

Use a soft cloth dampened with water and mild detergent to clean scale surfaces. 

 

Wipe surface with clean soft cloth dampened with water and then dry with clean soft cloth.  

 

Do not use abrasive materials to clean scale surface to prevent damage to the surface finish. 

 

Do not spray liquid directly onto scale surfaces.  Use only a damp cloth. 

DISINFECTION 

To disinfect the pediatric cradle and display / user interface: 

 

Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth.  Cloth 

cannot  be  dripping  wet.    Follow  manufacturer’s  instruction  on  the  proper  use  of 
commercially available disinfectants.  

 

Disinfectant  solutions  with  1%  sodium  hypochlorite  or  70%  isopropyl  alcohol  are  suitable 
for display / user interface and other scale contact surfaces. 

 

After disinfecting, use a soft cloth dampened with clean water and dry with a soft clean cloth 
to prevent buildup of material on scale finish.  

 

Do  not  use  abrasive  material  to  disinfect  /  clean  scale  surfaces  to  prevent  damage  to  the 
surface finish. 

 

Do not spray liquid directly onto scale surfaces.  Use only a damp cloth. 

 

WARNING:

 

DO  NOT  SPRAY  CLEANING  SOLUTION  OR  LIQUIDS  DIRECTLY 
ON SURFACES TO BE CLEANED

  

 

WARNING: 

EXPOSURE  TO  EXCESSIVE  LIQUID  WILL  DAMAGE  USER 
INTERFACE KEYPAD 

 
WARNING:

 

DO  NOT 

USE  PRESSURIZED  WATER  OR  STEAM.    THE  SCALE 

SYSTEM CONTAINS ELECTRONIC COMPONENTS THAT MAY BE 
ADVERSELY  AFFECTED  BY  EXPOSURE  TO  SUCH  AN 
ENVIRONMENT. 

 

Summary of Contents for SR Scales SR630i

Page 1: ...i KG_200915 Page 1 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by SI...

Page 2: ...S 7 SYSTEM DESCRIPTION AND INTENDED USE 8 MICROBAN ANTIMICROBIAL PROTECTION 8 STORAGE AND TRANSPORTATION 8 MAINTENANCE AND CLEANING 9 SPECIFICATIONS 10 BUTTON FUNCTIONS 11 BATTERY REPLACEMENT 13 THEOR...

Page 3: ...54 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc PACKING CHECKLIST Model SR630i KG Pediatric Scale DESCRIPTION QUANTITY SCALE BASE 1 ea AC POWER SUPPLY 1 ea RIGHT AN...

Page 4: ...QUANTITY SCALE BASE 1 ea AC POWER SUPPLY 1 ea RIGHT ANGLE ADAPTER FOR AC POWER SUPPLY 1 ea PACKAGE OF SIX 6 D CELL BATTERIES INSTALLED 6 ea CALIBRATION CERTIFICATE 1 ea WARRANTY CARD 1 ea MANUAL 1 ea...

Page 5: ...to acclimate to room temperature before use STEP 2 Figure 3 Install the six 6 D cell batteries into the Battery Compartment on underside of scale Attach the Battery Compartment Cover using the two 2...

Page 6: ...be STEP 4 Place one shelf over an inner support tube with a locking caster STEP 5 Place one 13 tube over the inner support tube STEP 6 Place one shelf over the inner support tube STEP 7 Place one 10 2...

Page 7: ...le top version STEP 13 Figure 5 Figure 3 Place the Scale on top of the cart aligning threaded holes on underside of scale with the holes on the Top Plate Ensure the Display is located to the back of t...

Page 8: ...s designed for moving the scale without a patient from one location to another This is not intended as a patient transport device Maximum weight capacity must not exceed 20 kilograms gross weight MICR...

Page 9: ...loth DISINFECTION To disinfect the pediatric cradle and display user interface Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow m...

Page 10: ...LL button to recall last stored displayed reading AVERAGING Automatic digital filter POWER SUPPLY Six 6 D cell batteries Medical Grade AC Wall Power Supply with adapter CALIBRATION Calibration is trac...

Page 11: ...l read 0 000 Kg This is used to zero the system before placing the patient on the scale system This action also resets previously stored weight values to zero Ensure that nothing is in contact with th...

Page 12: ...button to select digit To change digit use the DOWN arrow button Press ENTER to save changes NOTE When selected the year position defaults to 00 UP Press UP to increase the scale s on time or to sele...

Page 13: ...t is indicated the weight will be stored in memory NOTE If there are any accessories blanket diaper etc place them on the scale before zeroing the system This will ensure that the patient s NET weight...

Page 14: ...is amplified by the operational amplifier The amplifier is configured to current sum the output of each cell The output of the operational amplifier is digitized by the analog to digital converter Th...

Page 15: ...elect known calibrated weights traceable to NIST adding up to the maximum scale capacity NOTE DO NOT USE barbells or uncalibrated weights STEP 2 Zero the scale by pressing the ZERO button STEP 3 Place...

Page 16: ...ut the scale system into the Calibration Mode by switching the calibration switch on the display board Figure 10 CALIBRATION will flash on the display IMPORTANT CALIBRATION CHECK Qualified service per...

Page 17: ...6 Display will read ADD HALF Place ten kilograms of calibrated weight HALF the maximum scale capacity on the scale Allow weight to stabilize Press UP arrow to save change STEP 7 Display will read ADD...

Page 18: ...bilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no e...

Page 19: ...ting and Service Manual Part No MAN630i KG_200915 Page 19 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com...

Page 20: ...No MAN630i KG_200915 Page 20 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments...

Reviews: