CALIBRATION (Cont’d) 

 

STEP 3:

 

Select  at  least  two (2) known  calibrated weights  (traceable to  NIST) adding up to 

the maximum scale capacity. 

NOTE:

 

The first weight or group of weights should be half of the maximum scale capacity.  

The second weight or group of weights should add up to the maximum scale capacity.   

DO NOT USE

 barbells or uncalibrated weights. 

STEP 4:

  Press 

MENU

  button  until  “

3  Pt  Calibration

”  is  displayed.    Press  the 

UP

  arrow 

button. 

STEP  5:   

Display  will  read  “

ZERO

”.    Ensure  that  all  weight  has  been  removed  from  the 

platform and press the “

UP” 

arrow button. 

STEP 6:

  Display will read “

ADD HALF

”.  Place ten kilograms of calibrated weight, 

HALF

 

the maximum scale capacity on the scale.  Allow weight to stabilize.  Press 

UP

 arrow to save 

change.  

STEP  7:

    Display  will  read  “

ADD  FULL

”.    Place  twenty  kilograms  of  calibrated  weight, 

FULL 

scale of the maximum scale capacity on the scale.  Allow weight to stabilize.  Press 

UP 

arrow to save change. 

STEP 8

:  Press 

ENTER

 to save the calibration, or 

WEIGH 

to exit without saving

.

 

STEP 9:

  Switch the scale system out of Calibration Mode on the display board (Figure 10) 

and replace end cap. 

 

 
 

TROUBLESHOOTING 

 
 

SYMPTOM 

REASON/CORRECTIVE ACTION 

Low readings 

Check offset and verify calibration 

Ensure the cradle is not contacting the plate under the 
scale 

Only half of display activates 

Press the “WEIGH” button   

No power 

Check that batteries are making contact with each 
other and the terminals of the battery holder. 

Ensure AC power supply is firmly plugged in 

For additional information or assistance, telephone the Service Hotline: 1-800-654-6360 

or e-mail: 

[email protected]

 

Summary of Contents for SR Scales SR630i

Page 1: ...i KG_200915 Page 1 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by SI...

Page 2: ...S 7 SYSTEM DESCRIPTION AND INTENDED USE 8 MICROBAN ANTIMICROBIAL PROTECTION 8 STORAGE AND TRANSPORTATION 8 MAINTENANCE AND CLEANING 9 SPECIFICATIONS 10 BUTTON FUNCTIONS 11 BATTERY REPLACEMENT 13 THEOR...

Page 3: ...54 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc PACKING CHECKLIST Model SR630i KG Pediatric Scale DESCRIPTION QUANTITY SCALE BASE 1 ea AC POWER SUPPLY 1 ea RIGHT AN...

Page 4: ...QUANTITY SCALE BASE 1 ea AC POWER SUPPLY 1 ea RIGHT ANGLE ADAPTER FOR AC POWER SUPPLY 1 ea PACKAGE OF SIX 6 D CELL BATTERIES INSTALLED 6 ea CALIBRATION CERTIFICATE 1 ea WARRANTY CARD 1 ea MANUAL 1 ea...

Page 5: ...to acclimate to room temperature before use STEP 2 Figure 3 Install the six 6 D cell batteries into the Battery Compartment on underside of scale Attach the Battery Compartment Cover using the two 2...

Page 6: ...be STEP 4 Place one shelf over an inner support tube with a locking caster STEP 5 Place one 13 tube over the inner support tube STEP 6 Place one shelf over the inner support tube STEP 7 Place one 10 2...

Page 7: ...le top version STEP 13 Figure 5 Figure 3 Place the Scale on top of the cart aligning threaded holes on underside of scale with the holes on the Top Plate Ensure the Display is located to the back of t...

Page 8: ...s designed for moving the scale without a patient from one location to another This is not intended as a patient transport device Maximum weight capacity must not exceed 20 kilograms gross weight MICR...

Page 9: ...loth DISINFECTION To disinfect the pediatric cradle and display user interface Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow m...

Page 10: ...LL button to recall last stored displayed reading AVERAGING Automatic digital filter POWER SUPPLY Six 6 D cell batteries Medical Grade AC Wall Power Supply with adapter CALIBRATION Calibration is trac...

Page 11: ...l read 0 000 Kg This is used to zero the system before placing the patient on the scale system This action also resets previously stored weight values to zero Ensure that nothing is in contact with th...

Page 12: ...button to select digit To change digit use the DOWN arrow button Press ENTER to save changes NOTE When selected the year position defaults to 00 UP Press UP to increase the scale s on time or to sele...

Page 13: ...t is indicated the weight will be stored in memory NOTE If there are any accessories blanket diaper etc place them on the scale before zeroing the system This will ensure that the patient s NET weight...

Page 14: ...is amplified by the operational amplifier The amplifier is configured to current sum the output of each cell The output of the operational amplifier is digitized by the analog to digital converter Th...

Page 15: ...elect known calibrated weights traceable to NIST adding up to the maximum scale capacity NOTE DO NOT USE barbells or uncalibrated weights STEP 2 Zero the scale by pressing the ZERO button STEP 3 Place...

Page 16: ...ut the scale system into the Calibration Mode by switching the calibration switch on the display board Figure 10 CALIBRATION will flash on the display IMPORTANT CALIBRATION CHECK Qualified service per...

Page 17: ...6 Display will read ADD HALF Place ten kilograms of calibrated weight HALF the maximum scale capacity on the scale Allow weight to stabilize Press UP arrow to save change STEP 7 Display will read ADD...

Page 18: ...bilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no e...

Page 19: ...ting and Service Manual Part No MAN630i KG_200915 Page 19 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com...

Page 20: ...No MAN630i KG_200915 Page 20 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments...

Reviews: